Recall Record

Arlington, TN, United States

MicroPort Orthopedics Inc. recalls

7 recalls recorded since 2018, most recently on 19 December 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)4
Terminated or completed3
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
19 December 2025
Reason
Due to products not having FDA Premarket authorization to be distributed within the United States.
Distribution
The product was sent to distributors in multiple US states, including Texas, California, Wisconsin, Kansas, Illinois, Michigan, Tennessee, Maryland, Washington, Georgia, Colorado, Missouri, and Florida. The product was also sent to the following countries: Belgium, Germany, France, Italy, United Ki
Initiated
Voluntary: Firm initiated
Recall number
Z-1133-2026

Ongoing Class II Medical devices

Recall started
29 October 2024
Reason
Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.
Distribution
OUS: CN, GR, ES, ZA, FR, JP, IT
Initiated
Voluntary: Firm initiated
Recall number
Z-0699-2025

Ongoing Class II Medical devices

Recall started
3 May 2024
Reason
One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3 Evolution MP Non-Porous Femoral Left in the packaging.
Distribution
International distribution in the country of China. There is no field inventory in the United States for either lot.
Initiated
Voluntary: Firm initiated
Recall number
Z-2244-2024

Terminated Class II Medical devices

Recall started
28 October 2022
Terminated
10 September 2026
Reason
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
Distribution
International Distribution to countries of: Germany, Italy, France, Belgium, Austria, Greece, Romania, Japan, and P.R. China.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0276-2023, Z-0275-2023

Terminated Class II Medical devices

Recall started
4 May 2022
Terminated
16 January 2024
Reason
One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, IN, MO, CA and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1225-2022

Ongoing Class II Medical devices

Recall started
31 July 2020
Reason
MicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMUR¿ Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMUR¿ Hip System Package Insert with the most recent revision (150803-8).
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MO, MT, NC, NH, NJ, NV, OH, PA, TN, TX, UT, VA, WA, WI including PR; and countries of: Brazil, Canada, China, Costa Rica, France, Germany, Japan, The Netherlands and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2948-2020

Terminated Class I Medical devices

Recall started
31 July 2020
Terminated
16 April 2024
Reason
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR¿ Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
Distribution
worldwide, except China and Japan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2941-2020, Z-2942-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.