Micromed S.p.A. recalls

1 recall recorded since 2018, most recently on 4 May 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 May 2026
Reason
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
Distribution
Worldwide - US Nationwide distribution in the states of FL, PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2353-2026