Micro Therapeutics, Inc. recalls

7 recalls recorded since 2018, most recently on 25 June 2026. 7 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)7
Terminated or completed0
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
25 June 2026
Reason
Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.
Distribution
U.S. Nationwide distribution in the states of GA and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-3326-2026

Ongoing Class II Medical devices

Recall started
8 June 2026
Reason
Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
Distribution
US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2774-2026, Z-2773-2026

Ongoing Class I Medical devices

Recall started
30 January 2025
Reason
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
Distribution
Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Austra
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1246-2025, Z-1245-2025

Ongoing Class II Medical devices

Recall started
11 January 2024
Reason
European version of microcatheter were distributed within US which contain a different "Indications for Use".
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, MA, MD, MI, MO, NC, ND, NE, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1253-2024, Z-1252-2024

Ongoing Class II Medical devices

Recall started
30 June 2022
Reason
Outer carton packaging of microcatheters were exposed to wet substance during transport.
Distribution
US Nationwide Distribution: None OUS: Brazil
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1556-2022, Z-1557-2022

Ongoing Class II Medical devices

Recall started
27 April 2022
Reason
Due to incorrect size and configuration labeling of the detachable coil system.
Distribution
International distribution in the countries of China and Republic of Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-1261-2022

Ongoing Class I Medical devices

Recall started
13 July 2021
Reason
Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV O.U.S.: Argentina, Brazil Canada, Chile, Colombia, Ecuador, Mexico, Peru, Austria,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2403-2021, Z-2402-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.