Meridian Bioscience Inc recalls

9 recalls recorded since 2018, most recently on 5 December 2025. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)6
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
5 December 2025
Reason
The affected lots show a decline in performance over time, which may lead to false-negative results.
Distribution
US Nationwide distribution and the OUS country of Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-1019-2026

Ongoing Class II Medical devices

Recall started
7 October 2025
Reason
The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.
Distribution
US Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0568-2026, Z-0569-2026

Ongoing Class II Medical devices

Recall started
19 October 2023
Reason
Due to a manufacturing defect, certain Para-Pak vials may be at risk for leakage.
Distribution
Nationwide Foreign: Italy
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0521-2024, Z-0520-2024, Z-0519-2024

Ongoing Class II Medical devices

Recall started
15 December 2022
Reason
Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2
Distribution
US Nationwide distribution in the states of AZ, CA, IN, KS, KY, LA, MI, MN, NC, NY, OR, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1008-2023

Ongoing Class II Medical devices

Recall started
22 June 2022
Reason
When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Chile, Italy, Qatar, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1479-2022

Ongoing Class II Medical devices

Recall started
15 February 2022
Reason
The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OR, PA, PR, RI, SD, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Chile, Italy, Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0836-2022

Terminated Class II Medical devices

Recall started
8 April 2021
Terminated
26 October 2021
Reason
Emergency Use Application (EUA) application withdrawn as false negative results reported
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1540-2021, Z-1541-2021

Terminated Class II Medical devices

Recall started
8 February 2021
Terminated
26 October 2021
Reason
Use of the kits after incorrect storage could lead to a potential false-negative or invalid result. In the event of a false-negative, the incorrect assumption would be made as to the cause of patient symptoms. The appropriate treatment may not be administered, and symptoms may progress. Additional evaluations will be completed and documented within the associated Health Hazard Evaluation.
Distribution
US Nationwide distribution in the states of OH, SC.
Initiated
Voluntary: Firm initiated
Recall number
Z-1299-2021

Terminated Class II Medical devices

Recall started
17 December 2018
Terminated
30 April 2020
Reason
An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, CA, CO, DE, IL, KS, KY, LA, MD, MO, NC, ND, NJ, NY, SC, TX, and VA. The products were distributed to the following foreign countries: India and Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-0732-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.