Medtronic Sofamor Danek USA Inc recalls

21 recalls recorded since 2018, most recently on 12 August 2026. 9 are listed as ongoing. Status and classification are as published by the FDA.

Recalls21
Ongoing (FDA)9
Terminated or completed12
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
12 August 2026
Reason
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Distribution
US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0138-2027, Z-0137-2027

Ongoing Class II Medical devices

Recall started
16 September 2025
Reason
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Distribution
Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia,
Initiated
Voluntary: Firm initiated
Recall number
Z-0580-2026

Ongoing Class II Medical devices

Recall started
26 September 2023
Reason
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
Distribution
US, Colombia, S. Korea, New Zealand, India, Taiwan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0162-2024, Z-0163-2024, Z-0165-2024, Z-0166-2024, Z-0167-2024, Z-0164-2024, Z-0161-2024

Ongoing Class II Medical devices

Recall started
31 July 2023
Reason
The incorrect product labeling was applied to the product indicating the wrong size of product.
Distribution
US Nationwide distribution in the states of ID, IN, MN, NC, NY, OH, OR, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2524-2023

Ongoing Class II Medical devices

Recall started
12 July 2023
Reason
Incorrect Laser Marking - Solera Awl Tip Tap
Distribution
US Nationwide distribution in the states of FL, CO, IN, SC, TX, UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-2363-2023

Ongoing Class II Medical devices

Recall started
6 June 2023
Reason
There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
Distribution
US Nationwide distribution in the states of IN, WI, GA, OR, AR, OH, TN, TX, NC, OR, NE, MN, FL.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2123-2023, Z-2121-2023, Z-2122-2023, Z-2118-2023, Z-2117-2023, Z-2120-2023, Z-2119-2023

Ongoing Class II Medical devices

Recall started
16 February 2023
Reason
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1409-2023, Z-1412-2023, Z-1410-2023, Z-1413-2023, Z-1411-2023

Ongoing Class II Medical devices

Recall started
8 February 2023
Reason
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechte
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1331-2023, Z-1327-2023, Z-1324-2023, Z-1334-2023, Z-1333-2023, Z-1326-2023, Z-1329-2023, Z-1328-2023, Z-1323-2023, Z-1330-2023, Z-1332-2023, Z-1325-2023

Ongoing Class II Medical devices

Recall started
19 November 2021
Reason
Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.
Distribution
US: IN, CT, NE
Initiated
Voluntary: Firm initiated
Recall number
Z-1574-2023

Terminated Class II Medical devices

Recall started
2 September 2021
Terminated
5 October 2023
Reason
Engraving on implants may not match what is listed on the pouch labels.
Distribution
Distribution in US: CA, MN, MO, NC, NV, NY, OH, OK, OR, PA, SC, SD, WA, WV
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0086-2022, Z-0085-2022

Terminated Class II Medical devices

Recall started
27 April 2021
Terminated
20 August 2024
Reason
Product is impacted by a thread profile defect due to a manufacturing issue.
Distribution
International distribution in the countries of France, Switzerland and Jordan. No US distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1797-2021

Terminated Class II Medical devices

Recall started
17 March 2021
Terminated
3 May 2023
Reason
The shelf box identifies a small device actually contains a large device, and the shelf box identifying a large device actually contains a small device.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1440-2021, Z-1439-2021

Terminated Class II Medical devices

Recall started
8 June 2020
Terminated
10 May 2022
Reason
Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Belgium and Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2693-2020, Z-2694-2020, Z-2695-2020, Z-2692-2020, Z-2691-2020, Z-2690-2020

Terminated Class II Medical devices

Recall started
24 April 2020
Terminated
31 March 2021
Reason
Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
Distribution
Nationwide distribution to CA, FL, IL, IN, KY, MN, MO, MT, OH, PA, TX, VA, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2390-2020, Z-2389-2020, Z-2391-2020

Terminated Class II Medical devices

Recall started
26 March 2020
Terminated
26 March 2021
Reason
Nonconforming product; length of the product measured shorter than the labeled length.
Distribution
US Nationwide distribution in the states of CA and MI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1941-2020

Terminated Class II Medical devices

Recall started
27 September 2018
Terminated
16 September 2020
Reason
The Extra Small INFUSE Bone Graft kit from lot M111806AAA, may contain two absorbable collagen sponges (ACS), but should only contain one ACS.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-0440-2019

Terminated Class II Medical devices

Recall started
9 July 2018
Terminated
7 May 2021
Reason
This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-2603-2018

Terminated Class II Medical devices

Recall started
23 March 2018
Terminated
25 July 2021
Reason
Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.
Distribution
Worldwide Distribution - US Nationwide in the states of GA, CT, LA, MD, WY, IN, MD, FL, and France
Initiated
Voluntary: Firm initiated
Recall number
Z-1750-2018

Terminated Class II Medical devices

Recall started
13 March 2018
Terminated
2 November 2020
Reason
The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1641-2018

Terminated Class II Medical devices

Recall started
22 December 2017
Terminated
31 July 2018
Reason
Medtronic has discovered the product lot in question was noted as having a longer drill tip extension than what is indicated on the print specification.
Distribution
MN, NC
Initiated
Voluntary: Firm initiated
Recall number
Z-0566-2018

Terminated Class II Medical devices

Recall started
6 December 2017
Terminated
4 November 2019
Reason
Medtronic has discovered the directional arrow at the proximal end of the instrument may not correctly align with the cut out opening on the distal end of the instrument.
Distribution
US. Austria, Belgium, Bulgaria, Canada, France, Germany, Italy, Lebanon, Luxembourg, Portugal, South Africa, US
Initiated
Voluntary: Firm initiated
Recall number
Z-0352-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.