Medtronic Navigation, Inc. recalls
22 recalls recorded since 2018, most recently on 4 December 2024. 14 are listed as ongoing. Status and classification are as published by the FDA.
Recalls22
Ongoing (FDA)14
Terminated or completed8
Class I (FDA)7
Ongoing Class II Medical devices
- Recall started
- 4 December 2024
- Reason
- Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Aus
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1022-2025
- Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile Percutaneous Reference Pin Set is intended to be used to place and remove a pin percutaneo
Ongoing Class II Medical devices
- Recall started
- 22 August 2024
- Reason
- Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.
- Distribution
- Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0050-2025
- StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or per
Ongoing Class I Medical devices
- Recall started
- 26 March 2024
- Reason
- Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Gua
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1776-2024
- Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.
Ongoing Class II Medical devices
- Recall started
- 28 November 2023
- Reason
- Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Oman, Portugal, Spain, Switzerl
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0698-2024, Z-0699-2024
- Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUS
- Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS
Ongoing Class I Medical devices
- Recall started
- 27 September 2023
- Reason
- Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.
- Distribution
- Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.(foreign) including countries of
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0188-2024
- StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneo
Ongoing Class I Medical devices
- Recall started
- 11 April 2023
- Reason
- During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
- Distribution
- Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIU
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1630-2023
- StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
Ongoing Class I Medical devices
- Recall started
- 11 April 2023
- Reason
- In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
- Distribution
- US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREEC
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1666-2023
- StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
Ongoing Class II Medical devices
- Recall started
- 15 December 2022
- Reason
- Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery.
- Distribution
- Worldwide - U.S. distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The country of Japan.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1039-2023
- AxiEM" Non-Invasive Patient Tracker
Ongoing Class II Medical devices
- Recall started
- 8 December 2022
- Reason
- Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.
- Distribution
- US Nationwide distribution including in the states of OH, CA, FL, NJ, NY, MN, IN, VA, WA, MS, SC, MO, GA, PA, KS, TN, LA, TX, ME, MA, AL, NC, MD, NH, IL, AZ, MI, NV, MT, OK, SD, RI, CO, CT, NM, NE, DE, WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0953-2023
- RadiaLux Lighted Retractor (Pink), REF: 50-101-1
Ongoing Class II Medical devices
- Recall started
- 14 October 2022
- Reason
- Due to a defect in the outer pouch sterile seal
- Distribution
- U.S. Nationwide distribution including in the states AZ, CA, CO, DC, FL, IL, KY, MA, MD, ME, MI, MN, MO, NH, NJ, NY, OH, PA, RI, TX, VA, WA, and WV.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0244-2023
- Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulate soft tissue by physicians in minimally invasive surgical procedures.
Ongoing Class I Medical devices
- Recall started
- 11 November 2021
- Reason
- Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.
- Distribution
- US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS, WY, MN, PR OUS: AU, BN, GU, JP, KR, MY, NZ, SG, TH, VN, MM, CA, CZ, RU, AL, CY, DE, ES, FR, GB, HU, IL,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0383-2022
- Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585
Ongoing Class II Medical devices
- Recall started
- 9 November 2021
- Reason
- A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
- Distribution
- US: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT, MS, VA, SC. OUS: Canada, Sweden, Italy, France, Germany, Spain, Czech Republic, United Kingdom, India, Nepal, Japan, Turkey, China
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0417-2022
- Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.
Terminated Class II Medical devices
- Recall started
- 5 June 2020
- Terminated
- 28 May 2024
- Reason
- There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.
- Distribution
- US Nationwide distribution including in the states of Alabama and Massachusetts.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2707-2020
- StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
Ongoing Class II Medical devices
- Recall started
- 5 December 2019
- Reason
- There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]
- Distribution
- US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2330-2020
- Mazor X Surgical System Positioner Type II, REF: ASM0214-02
Ongoing Class II Medical devices
- Recall started
- 30 October 2019
- Reason
- The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
- Distribution
- US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jerse
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0658-2020, Z-0659-2020, Z-0657-2020
- Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517
- Biopsy Needle FPU Kit, Product # 9736075, UDI: 00763000063108
- Passive Biopsy Needle Kit, UDI: 00643169030121 and 00643169702943
Terminated Class II Medical devices
- Recall started
- 28 August 2019
- Terminated
- 5 May 2021
- Reason
- Product was incorrectly assembled which could affect navigation accuracy of the device.
- Distribution
- U.S.: NY No O.U.S.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1693-2020
- NavLock Tracker *** Rx Only
Terminated Class I Medical devices
- Recall started
- 9 August 2019
- Terminated
- 2 April 2024
- Reason
- Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
- Distribution
- Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1995-2020
- Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing t
Terminated Class II Medical devices
- Recall started
- 8 July 2019
- Terminated
- 16 October 2024
- Reason
- The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
- Distribution
- US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1905-2020
- Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter
Terminated Class II Medical devices
- Recall started
- 6 December 2018
- Terminated
- 17 December 2020
- Reason
- Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.
- Distribution
- Distribution was nationwide. There was government/military/foreign distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0707-2020
- Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF
Terminated Class I Medical devices
- Recall started
- 26 September 2018
- Terminated
- 5 September 2024
- Reason
- An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
- Distribution
- US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0552-2019
- StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx
Terminated Class II Medical devices
- Recall started
- 1 June 2018
- Terminated
- 2 February 2022
- Reason
- Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.
- Distribution
- U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland, Netherlands, Israel
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-3251-2018
- Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542
Terminated Class II Medical devices
- Recall started
- 23 March 2018
- Terminated
- 31 July 2023
- Reason
- Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International distribution to Israel.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2501-2018
- Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase¿ Cooled Laser Applicator System 10mm Tip;
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.