Medtronic Navigation, Inc.-Littleton recalls

8 recalls recorded since 2018, most recently on 26 April 2024. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)2
Terminated or completed6
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
26 April 2024
Reason
Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, MN, PR, TX, WI and the countries of Bangladesh, Canada, Costa Rica, India, Mexico, Panama, Philippines.
Initiated
Voluntary: Firm initiated
Recall number
Z-2026-2024

Ongoing Class II Medical devices

Recall started
25 July 2023
Reason
Ground cable installed incorrectly.
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KY, MI, MO, NC, NV, OH, TN, UT, VA, & WA. The countries of Canada, Ireland, Poland, Russian Federation, Switzerland, & Viet Nam.
Initiated
Voluntary: Firm initiated
Recall number
Z-2449-2023

Terminated Class II Medical devices

Recall started
15 November 2019
Terminated
19 September 2023
Reason
Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.
Distribution
US: AR CA CO CT FL ID IL IN KS KY MI MN MO NC ND NE NY OH OR PA SC SD TX VA WA WI WV OUS: AUSTRALIA CANADA CHINA FRANCE GERMANY INDIA ITALY JAPAN LEBANON NETHERLANDS ROMANIA SAUDI ARABIA SPAIN SWEDEN SWITZERLAND UNITED KINGDOM
Initiated
Voluntary: Firm initiated
Recall number
Z-0767-2020

Terminated Class II Medical devices

Recall started
1 May 2019
Terminated
19 January 2021
Reason
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
Distribution
Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN K
Initiated
Voluntary: Firm initiated
Recall number
Z-1932-2019

Terminated Class II Medical devices

Recall started
17 September 2018
Terminated
28 May 2020
Reason
O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit
Distribution
Worldwide Distribution - US Nationwide Foreign: AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BRAZIL CANADA CHILE COLOMBIA DENMARK FINLAND FRANCE GERMANY HR ICELAND INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KOREA, REPUBLIC OF LUXEMBOURG MARTINIQUE MEXICO NETHERLANDS NEW ZEAL
Initiated
Voluntary: Firm initiated
Recall number
Z-0404-2019

Terminated Class II Medical devices

Recall started
30 August 2018
Terminated
18 June 2020
Reason
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
Distribution
Worldwide - US Distribution in the states: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and Puerto Rico. The products were distributed to the following foreign countr
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0271-2019, Z-0272-2019, Z-0280-2019, Z-0278-2019, Z-0276-2019, Z-0275-2019, Z-0274-2019, Z-0273-2019, Z-0277-2019, Z-0279-2019

Terminated Class II Medical devices

Recall started
25 January 2018
Terminated
1 April 2019
Reason
Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report, Gantry motion, Network communication, System shutdown, and Early termination of 3D spin.
Distribution
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Puerto Rico. Distributed in Algeria, Australia, Chile, China, Czech Republic, F
Initiated
Voluntary: Firm initiated
Recall number
Z-0928-2018

Terminated Class II Medical devices

Recall started
30 September 2017
Terminated
2 May 2019
Reason
During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnecting process which may result in an electrical short and a shock to the field service technician.
Distribution
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV., and to the countries of : Australia, Austria
Initiated
Voluntary: Firm initiated
Recall number
Z-0927-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.