Recall Record

Northfield, IL, United States

Medline Industries, LP recalls

118 recalls recorded since 2018, most recently on 11 September 2026. 83 are listed as ongoing. Status and classification are as published by the FDA.

Recalls118
Ongoing (FDA)83
Terminated or completed35
Class I (FDA)17

Ongoing Class II Medical devices

Recall started
11 September 2026
Reason
Discrepancy in the shelf-life of a component
Distribution
US Nationwide distribution in the state of NY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0189-2027

Ongoing Class II Medical devices

Recall started
2 September 2026
Reason
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0176-2027, Z-0178-2027, Z-0172-2027, Z-0177-2027, Z-0170-2027, Z-0174-2027, Z-0175-2027, Z-0179-2027, Z-0171-2027, Z-0169-2027, Z-0168-2027, Z-0173-2027

Ongoing Class II Medical devices

Recall started
31 August 2026
Reason
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Distribution
US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0184-2027, Z-0183-2027, Z-0185-2027, Z-0182-2027, Z-0181-2027, Z-0180-2027

Ongoing Class I Medical devices

Recall started
14 August 2026
Reason
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Distribution
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3228-2026, Z-3227-2026

Ongoing Class II Medical devices

Recall started
14 August 2026
Reason
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0087-2027, Z-0086-2027

Ongoing Class I Medical devices

Recall started
11 August 2026
Reason
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3125-2026, Z-3121-2026, Z-3126-2026, Z-3122-2026, Z-3127-2026, Z-3124-2026, Z-3123-2026

Ongoing Class II Medical devices

Recall started
10 August 2026
Reason
Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of the brush. Potential hazards associated include cut injury/abrasion and infection.
Distribution
US Nationwide distribution in the states of MN and SD.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3229-2026, Z-3230-2026

Ongoing Class II Medical devices

Recall started
10 August 2026
Reason
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3269-2026, Z-3268-2026, Z-3267-2026, Z-3265-2026, Z-3266-2026

Ongoing Class II Medical devices

Recall started
31 July 2026
Reason
Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.
Distribution
US Nationwide distribution in states of Arkansas, Florida, North Carolina, Maryland, and Texas.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3209-2026, Z-3207-2026, Z-3208-2026, Z-3210-2026

Ongoing Class II Medical devices

Recall started
23 July 2026
Reason
The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.
Distribution
US Nationwide distribution in the state of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-3137-2026

Ongoing Class II Medical devices

Recall started
20 July 2026
Reason
Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that were recalled by the manufacturer because they may have an open or incomplete seal on the applicator packaging. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3073-2026, Z-3078-2026, Z-3068-2026, Z-3070-2026, Z-3071-2026, Z-3075-2026, Z-3069-2026, Z-3074-2026, Z-3072-2026, Z-3079-2026, Z-3077-2026, Z-3076-2026 and 1 more

Ongoing Class II Medical devices

Recall started
13 July 2026
Reason
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Distribution
Worldwide distribution - US Nationwide and the country of Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2992-2026, Z-2991-2026, Z-2993-2026

Ongoing Class II Medical devices

Recall started
10 July 2026
Reason
Kits contain a potentially contaminated component which were recalled by the manufacturer
Distribution
US Nationwide distribution in the states of AL, CA, FL, IL, KS, LA, MA, MD, MO, MS, MT, NV, NY, OH, OK, SC, TX, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3029-2026, Z-3028-2026, Z-3022-2026, Z-3026-2026, Z-3023-2026, Z-3024-2026, Z-3019-2026, Z-3025-2026, Z-3020-2026, Z-3021-2026, Z-3027-2026

Ongoing Class II Medical devices

Recall started
10 July 2026
Reason
Component labeling discrepancy
Distribution
US Nationwide distribution in the state of California.
Initiated
Voluntary: Firm initiated
Recall number
Z-2999-2026

Ongoing Class II Medical devices

Recall started
8 July 2026
Reason
Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.
Distribution
US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2994-2026, Z-2996-2026, Z-2995-2026, Z-2997-2026

Ongoing Class II Medical devices

Recall started
30 June 2026
Reason
Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
Distribution
US Nationwide distribution in the states and territories of Guam, Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2926-2026, Z-2925-2026, Z-2924-2026, Z-2923-2026

Ongoing Class II Medical devices

Recall started
12 June 2026
Reason
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3057-2026, Z-3050-2026, Z-3053-2026, Z-3059-2026, Z-3060-2026, Z-3056-2026, Z-3061-2026, Z-3054-2026, Z-3049-2026, Z-3051-2026, Z-3048-2026, Z-3058-2026 and 4 more

Ongoing Class II Medical devices

Recall started
11 June 2026
Reason
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Distribution
US: CT, FL, KS, OH, OR, NY, MD, MT, VA, MA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2789-2026, Z-2791-2026, Z-2788-2026, Z-2790-2026

Ongoing Class II Medical devices

Recall started
10 June 2026
Reason
Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.
Distribution
US Nationwide distribution in the states of CA and UT.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2736-2026, Z-2737-2026

Ongoing Class I Medical devices

Recall started
8 June 2026
Reason
The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
Distribution
US: Mo, CO, IL
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2765-2026, Z-2764-2026

Ongoing Class II Medical devices

Recall started
8 June 2026
Reason
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Distribution
US Nationwide distribution in the states of CA, FL, MD. PA, and TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2784-2026, Z-2781-2026, Z-2779-2026, Z-2787-2026, Z-2782-2026, Z-2780-2026, Z-2785-2026, Z-2786-2026, Z-2783-2026

Ongoing Class I Medical devices

Recall started
5 June 2026
Reason
Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. Huons Co, Ltd. has issued a voluntary recall of this product due to quality issues.
Distribution
US: OH, CA, IL, IN, FL, SC, AZ, NJ.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2747-2026, Z-2748-2026, Z-2749-2026

Ongoing Class II Medical devices

Recall started
28 May 2026
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2609-2026, Z-2607-2026, Z-2611-2026, Z-2608-2026, Z-2612-2026, Z-2613-2026, Z-2606-2026, Z-2610-2026

Ongoing Class II Medical devices

Recall started
28 May 2026
Reason
Convenience Kits contain Covidien Mon-a-Therm General Purpose Temperature Probes Size 9 Fr. Medtronic has issued a recall of specific production lots of the Covidien Mon-a-Therm general purpose temperature probe-size 9 due to an error in its distribution control process.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2695-2026, Z-2697-2026, Z-2696-2026

Ongoing Class II Medical devices

Recall started
21 May 2026
Reason
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2542-2026, Z-2538-2026, Z-2539-2026, Z-2541-2026, Z-2540-2026, Z-2537-2026, Z-2534-2026, Z-2533-2026, Z-2544-2026, Z-2535-2026, Z-2547-2026, Z-2543-2026 and 3 more

Ongoing Class I Medical devices

Recall started
15 May 2026
Reason
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Distribution
US Nationwide distribution in the state of TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2615-2026

Ongoing Class I Medical devices

Recall started
6 May 2026
Reason
The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2452-2026, Z-2455-2026, Z-2454-2026, Z-2453-2026

Ongoing Class II Medical devices

Recall started
4 May 2026
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2486-2026, Z-2496-2026, Z-2487-2026, Z-2485-2026, Z-2502-2026, Z-2489-2026, Z-2505-2026, Z-2484-2026, Z-2490-2026, Z-2492-2026, Z-2493-2026, Z-2499-2026 and 18 more

Ongoing Class II Medical devices

Recall started
27 April 2026
Reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Distribution
US Nationwide distribution. OUS distribution pending.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2438-2026, Z-2434-2026, Z-2436-2026, Z-2431-2026, Z-2416-2026, Z-2433-2026, Z-2430-2026, Z-2427-2026, Z-2417-2026, Z-2429-2026, Z-2414-2026, Z-2432-2026 and 22 more

Ongoing Class I Medical devices

Recall started
10 April 2026
Reason
Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Distribution
Worldwide - US Nationwide distribution in the US Territory of Virgin Islands and the countries of Bahamas, Panama, Barbados.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2234-2026, Z-2236-2026, Z-2235-2026, Z-2237-2026, Z-2233-2026, Z-2231-2026, Z-2232-2026

Ongoing Class II Medical devices

Recall started
10 April 2026
Reason
Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.
Distribution
US Nationwide distribution in the states of California.
Initiated
Voluntary: Firm initiated
Recall number
Z-2260-2026

Ongoing Class I Medical devices

Recall started
24 March 2026
Reason
Medline has identified the presence of particulate within the fluid path of the Manifolds.
Distribution
US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2146-2026, Z-2141-2026, Z-2148-2026, Z-2143-2026, Z-2145-2026, Z-2144-2026, Z-2139-2026, Z-2138-2026, Z-2140-2026, Z-2142-2026, Z-2147-2026

Ongoing Class II Medical devices

Recall started
19 March 2026
Reason
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2008-2026, Z-2037-2026, Z-2012-2026, Z-2014-2026, Z-2030-2026, Z-2035-2026, Z-2023-2026, Z-2022-2026, Z-2011-2026, Z-2029-2026, Z-2027-2026, Z-2038-2026 and 19 more

Ongoing Class II Medical devices

Recall started
16 March 2026
Reason
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2040-2026, Z-2045-2026, Z-2041-2026, Z-2043-2026, Z-2042-2026, Z-2039-2026, Z-2044-2026

Ongoing Class II Medical devices

Recall started
16 March 2026
Reason
On lot of Sterile Circumcision Kits was distributed without the required sterilization process and is therefore not sterile.
Distribution
US Nationwide distribution in the state of South Carolina.
Initiated
Voluntary: Firm initiated
Recall number
Z-1824-2026

Ongoing Class II Medical devices

Recall started
27 February 2026
Reason
Unapproved design changes to the products outside of the 510(k) clearance.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1885-2026, Z-1878-2026, Z-1884-2026, Z-1883-2026, Z-1876-2026, Z-1886-2026, Z-1892-2026, Z-1888-2026, Z-1887-2026, Z-1877-2026, Z-1882-2026, Z-1889-2026 and 5 more

Ongoing Class I Medical devices

Recall started
27 February 2026
Reason
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1719-2026, Z-1713-2026, Z-1716-2026, Z-1714-2026, Z-1720-2026, Z-1728-2026, Z-1712-2026, Z-1718-2026, Z-1723-2026, Z-1722-2026, Z-1715-2026, Z-1729-2026 and 7 more

Ongoing Class II Medical devices

Recall started
27 February 2026
Reason
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2590-2026, Z-2591-2026, Z-2587-2026, Z-2582-2026, Z-2592-2026, Z-2585-2026, Z-2581-2026, Z-2594-2026, Z-2588-2026, Z-2595-2026, Z-2584-2026, Z-2589-2026 and 4 more

Ongoing Class II Medical devices

Recall started
25 February 2026
Reason
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1838-2026, Z-1829-2026, Z-1835-2026, Z-1834-2026, Z-1847-2026, Z-1836-2026, Z-1839-2026, Z-1830-2026, Z-1841-2026, Z-1845-2026, Z-1833-2026, Z-1840-2026 and 10 more

Ongoing Class II Medical devices

Recall started
18 February 2026
Reason
There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.
Distribution
US Nationwide distribution in the state of MN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1693-2026

Ongoing Class II Medical devices

Recall started
16 February 2026
Reason
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1707-2026, Z-1705-2026, Z-1709-2026, Z-1710-2026, Z-1706-2026, Z-1708-2026, Z-1711-2026

Ongoing Class II Medical devices

Recall started
13 February 2026
Reason
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Distribution
Worldwide - US Nationwide and the countries of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1691-2026, Z-1690-2026

Ongoing Class II Medical devices

Recall started
10 February 2026
Reason
Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
Distribution
Domestic distribution nationwide US. International distribution to Mexico and Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-1650-2026

Ongoing Class II Medical devices

Recall started
30 January 2026
Reason
The instructions for use (IFUs) on specified Cardinal Health Chest Drainage Units (CDUs) and accessories have been updated to clarify the intended target patient population as adults, 18 years and older.
Distribution
Worldwide - US Nationwide distribution in the state of CA and the countries of Chile, Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-1562-2026

Ongoing Class II Medical devices

Recall started
28 January 2026
Reason
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Distribution
Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1563-2026, Z-1564-2026

Ongoing Class II Medical devices

Recall started
8 January 2026
Reason
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1464-2026, Z-1468-2026, Z-1465-2026, Z-1463-2026, Z-1467-2026, Z-1466-2026

Ongoing Class II Medical devices

Recall started
7 January 2026
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2099-2026, Z-2119-2026, Z-2104-2026, Z-2100-2026, Z-2093-2026, Z-2116-2026, Z-2090-2026, Z-2096-2026, Z-2095-2026, Z-2092-2026, Z-2115-2026, Z-2109-2026 and 30 more

Ongoing Class II Medical devices

Recall started
7 January 2026
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1392-2026, Z-1407-2026, Z-1401-2026, Z-1406-2026, Z-1402-2026, Z-1410-2026, Z-1419-2026, Z-1395-2026, Z-1405-2026, Z-1385-2026, Z-1394-2026, Z-1398-2026 and 30 more

Ongoing Class II Medical devices

Recall started
5 January 2026
Reason
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Distribution
Worldwide distribution - US Nationwide and the countries of Bahamas, Bermuda, Cayman Islands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1434-2026, Z-1435-2026, Z-1431-2026, Z-1433-2026, Z-1432-2026

Ongoing Class II Medical devices

Recall started
31 December 2025
Reason
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
Distribution
Worldwide distribution - US Nationwide and the country of Bahamas.
Initiated
Voluntary: Firm initiated
Recall number
Z-1366-2026

Ongoing Class II Medical devices

Recall started
24 December 2025
Reason
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1331-2026, Z-1328-2026, Z-1332-2026, Z-1326-2026, Z-1334-2026, Z-1335-2026, Z-1329-2026, Z-1330-2026, Z-1333-2026, Z-1327-2026, Z-1336-2026

Ongoing Class II Medical devices

Recall started
24 December 2025
Reason
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Distribution
Distribution US nationwide and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1312-2026, Z-1313-2026

Ongoing Class I Medical devices

Recall started
22 December 2025
Reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Distribution
Distribution US nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1320-2026, Z-1319-2026, Z-1325-2026, Z-1322-2026, Z-1324-2026, Z-1321-2026, Z-1323-2026

Ongoing Class II Medical devices

Recall started
16 December 2025
Reason
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1090-2026, Z-1089-2026, Z-1086-2026, Z-1085-2026, Z-1087-2026, Z-1088-2026

Ongoing Class II Medical devices

Recall started
12 December 2025
Reason
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1200-2026, Z-1186-2026, Z-1189-2026, Z-1191-2026, Z-1195-2026, Z-1194-2026, Z-1193-2026, Z-1199-2026, Z-1187-2026, Z-1196-2026, Z-1197-2026, Z-1188-2026 and 3 more

Ongoing Class II Medical devices

Recall started
26 November 2025
Reason
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1054-2026, Z-1055-2026, Z-1057-2026, Z-1051-2026, Z-1049-2026, Z-1058-2026, Z-1060-2026, Z-1059-2026, Z-1056-2026, Z-1053-2026, Z-1052-2026, Z-1050-2026

Ongoing Class I Medical devices

Recall started
26 November 2025
Reason
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Distribution
US, CA, PA, GUAM, VI
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1015-2026, Z-1014-2026, Z-1013-2026, Z-1016-2026, Z-1012-2026, Z-1011-2026

Ongoing Class II Medical devices

Recall started
21 November 2025
Reason
Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bahamas, Canada, China, Dominican Republic, Guam, South Korea, Mexico, Panama, and Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-1035-2026

Ongoing Class II Medical devices

Recall started
13 November 2025
Reason
The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.
Distribution
US Nationwide distribution in the states CA, CO, IL, KS, and MI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0921-2026, Z-0922-2026

Ongoing Class II Medical devices

Recall started
11 November 2025
Reason
There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CO, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MT, NC, NJ, OH, OK, PA, SC, TX, VA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0911-2026

Ongoing Class II Medical devices

Recall started
27 October 2025
Reason
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Distribution
US, Guam, Canada, United Arab Emirates
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0904-2026, Z-0903-2026, Z-0900-2026, Z-0902-2026, Z-0898-2026, Z-0899-2026, Z-0901-2026

Ongoing Class II Medical devices

Recall started
27 October 2025
Reason
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
Distribution
US Nationwide distribution in the state of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0886-2026

Ongoing Class I Medical devices

Recall started
10 October 2025
Reason
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Distribution
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0444-2026, Z-0442-2026, Z-0446-2026, Z-0445-2026, Z-0447-2026, Z-0451-2026, Z-0448-2026, Z-0443-2026, Z-0450-2026, Z-0449-2026

Ongoing Class II Medical devices

Recall started
3 October 2025
Reason
Medline has identified specific sterile eye procedure kits with pouch packaging that may exhibit open seals.
Distribution
US: MO
Initiated
Voluntary: Firm initiated
Recall number
Z-0510-2026

Ongoing Class II Medical devices

Recall started
1 October 2025
Reason
Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0471-2026, Z-0469-2026, Z-0470-2026, Z-0474-2026, Z-0477-2026, Z-0473-2026, Z-0472-2026, Z-0467-2026, Z-0468-2026, Z-0476-2026, Z-0475-2026

Ongoing Class II Medical devices

Recall started
30 September 2025
Reason
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0519-2026, Z-0525-2026, Z-0529-2026, Z-0514-2026, Z-0528-2026, Z-0524-2026, Z-0523-2026, Z-0526-2026, Z-0512-2026, Z-0511-2026, Z-0517-2026, Z-0513-2026 and 8 more

Ongoing Class II Drugs

Recall started
29 September 2025
Reason
Subpotent Drug
Distribution
Nationwide within the United States
Initiated
Voluntary: Firm initiated
Recall number
D-0035-2026

Ongoing Class II Medical devices

Recall started
11 September 2025
Reason
Medline has identified a limited quantity of saline IV flush syringes that may not meet the required concentration of sodium chloride.
Distribution
Distribution US nationwide and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-0234-2026

Ongoing Class II Medical devices

Recall started
10 September 2025
Reason
Specific items and lots of Medline Kits may contain Clearlink IV Sets which Baxter recalled due to customer reports of leaking.
Distribution
US Nationwide distribution in the states of MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-0137-2026

Ongoing Class I Medical devices

Recall started
21 August 2025
Reason
The kits contain certain lots of cannula products where the catheter may not retain its shape.
Distribution
US Nationwide distribution in the state of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2670-2025

Ongoing Class III Drugs

Recall started
19 August 2025
Reason
Subpotent drug
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0674-2025

Ongoing Class I Medical devices

Recall started
8 August 2025
Reason
Convenience kits labeled as sterile have not gone through the sterilization process.
Distribution
US Distribution to States of: FL, MD, NJ, TN
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2593-2025, Z-2592-2025

Ongoing Class I Medical devices

Recall started
31 July 2025
Reason
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Distribution
Domestic US distribution nationwide. International distribution pending. No international distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2610-2025, Z-2611-2025, Z-2612-2025, Z-2614-2025, Z-2613-2025

Ongoing Class II Medical devices

Recall started
21 July 2022
Reason
Due to a non-sterile component (tape) being place within the sterile portion of the Chest Tube Convenience Kit
Distribution
U.S. Nationwide distribution in the state of AL.
Initiated
Voluntary: Firm initiated
Recall number
Z-1683-2022

Terminated Class II Drugs

Recall started
8 November 2021
Terminated
28 August 2024
Reason
Defective Container: Customer complaints for leaking bottles and dispensing issues.
Distribution
Nationwide in the USA, Canada, Hong Kong, and Peru
Initiated
Voluntary: Firm initiated
Recall number
D-0287-2022

Terminated Class II Medical devices

Recall started
27 October 2021
Terminated
5 April 2023
Reason
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0358-2022, Z-0361-2022, Z-0360-2022, Z-0362-2022, Z-0359-2022

Terminated Class II Medical devices

Recall started
6 August 2021
Terminated
25 August 2025
Reason
Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Australia, Japan, Latvia, Lebanon, Malaysia, Netherlands, Pakistan, People's Republic of China, Panama, Singapore, South Africa, South Korea, Thailand, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2575-2021, Z-2574-2021

Terminated Class II Medical devices

Recall started
1 July 2021
Terminated
2 March 2023
Reason
The kits may contain an expired component.
Distribution
US Nationwide distribution in the states of MN, NE, and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2234-2021, Z-2232-2021, Z-2233-2021, Z-2231-2021

Terminated Class II Medical devices

Recall started
25 May 2021
Terminated
27 February 2023
Reason
Potential for mold contamination (Aspergillus vadensis)
Distribution
Nationwide Distribution to states of: AL, AR, AZ, CA, CO, DC, FL, IA, IL, IN, MD, MI, MN, MO, NJ, NY, OH, PA, RI, SC, TN, TX, and VA; International to country of: Antigua.
Initiated
Voluntary: Firm initiated
Recall number
Z-2080-2021

Terminated Class III Drugs

Recall started
14 May 2021
Terminated
7 August 2024
Reason
Superpotent Drug: Product is above specification for active ingredient, 2% Chlorhexidine Gluconate.
Distribution
Distributed Nationwide in the USA and Isreal
Initiated
Voluntary: Firm initiated
Recall number
D-0624-2021

Ongoing Class II Medical devices

Recall started
30 April 2021
Reason
The devices may not meet the minimum required sterility assurance level.
Distribution
US Distribution to state of: WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1841-2021

Ongoing Class II Medical devices

Recall started
15 March 2021
Reason
Possible false negative or false positive results due to the product being compromised during shipment.
Distribution
US Nationwide distribution in the state of Ohio.
Initiated
Voluntary: Firm initiated
Recall number
Z-1407-2021

Terminated Class II Medical devices

Recall started
25 January 2021
Terminated
9 January 2023
Reason
Product was compromised during shipment.
Distribution
US Nationwide distribution in the states of KY, PA, OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-1249-2021

Terminated Class III Medical devices

Recall started
15 January 2021
Terminated
9 January 2023
Reason
Medline Industries, Inc. is removing specific lots of STAT-Site¿ Controls (Low and High) S-O303000, from the market, due to the product being compromised during shipment.
Distribution
US Nationwide distribution in the states of MS, OH, IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-1135-2021

Terminated Class II Medical devices

Recall started
13 November 2020
Terminated
26 October 2021
Reason
The cylindrical sponge component is not x-ray detectable.
Distribution
Foreign distribution only - Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0779-2021, Z-0783-2021, Z-0782-2021, Z-0781-2021, Z-0780-2021

Terminated Class II Medical devices

Recall started
5 November 2020
Terminated
4 August 2021
Reason
RevMedX Trauma Dressing is being recalled due to package seal integrity.
Distribution
Worldwide distribution - US Nationwide distributions and the countries of Singapore, Denmark, Poland, Portugal, Kenya, Israel, and France.
Initiated
Voluntary: Firm initiated
Recall number
Z-0615-2021

Terminated Class II Medical devices

Recall started
1 October 2020
Terminated
3 May 2021
Reason
Product compromised during shipment.
Distribution
Domestic Distribution only to: CA, NY, TX and VA
Initiated
Voluntary: Firm initiated
Recall number
Z-0297-2021

Ongoing Class II Medical devices

Recall started
15 September 2020
Reason
The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation.
Distribution
Domestic Distribution Only: IL, MI
Initiated
Voluntary: Firm initiated
Recall number
Z-0055-2021

Terminated Class II Medical devices

Recall started
14 September 2020
Terminated
9 April 2021
Reason
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities.
Distribution
US Nationwide distribution including in the states of CA, CO, CT, OH, TX, and NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0062-2021, Z-0061-2021, Z-0066-2021, Z-0067-2021, Z-0064-2021, Z-0063-2021, Z-0065-2021, Z-0068-2021

Terminated Class II Medical devices

Recall started
1 September 2020
Terminated
27 May 2022
Reason
There have been reports of the suction catheter coming apart from the device assembly during use.
Distribution
62 US consignees; 1 government consignee, 4 foreign consignees
Initiated
Voluntary: Firm initiated
Recall number
Z-0058-2021

Terminated Class II Medical devices

Recall started
12 August 2020
Terminated
28 April 2021
Reason
Potential of incomplete seals impacting the sterility of the kit.
Distribution
Domestic Distribution: MN and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-3013-2020

Terminated Class II Medical devices

Recall started
28 July 2020
Terminated
13 February 2023
Reason
Product was shipped with refrigeration (ice) instead of frozen (dry ice).
Distribution
Distribution in US - AR, AZ, CO, FL, MD, MS, MI, NC, NM, NY, OK, PA, RI, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-3054-2020

Terminated Class II Medical devices

Recall started
9 July 2020
Terminated
25 January 2021
Reason
Potential risk of brush detachment during cleaning.
Distribution
US Nationwide distribution including in the states of CA, IN and OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-2766-2020

Ongoing Class II Medical devices

Recall started
30 June 2020
Reason
There are reports of the suction catheter coming apart from the device assembly during use.
Distribution
US Nationwide distribution including in the states of AR, AZ, CA, FL, GA, IL, MS, NY, NV, OH, PA, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2725-2020, Z-2724-2020

Terminated Class II Medical devices

Recall started
9 March 2020
Terminated
22 February 2022
Reason
Medline identified that there may be a potential for voids in the packaging seal and loss of sterility on the outside of the foil pouch.
Distribution
nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2040-2020

Terminated Class II Medical devices

Recall started
10 February 2020
Terminated
13 February 2023
Reason
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
Distribution
US Nationwide: MA, LA, IL, MO, MI, MN, NJ, KS, GA, NY, PA, SD, CT, TN, MD, OH, VA, FL, TX, KY, CO, IN, AL, CA, MT, OR, AK, AL, DC, IA, ME, NV, NE, SC, WV, WY, RI, NC, AZ, UT, DE and WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1544-2020, Z-1546-2020, Z-1545-2020

Terminated Class II Medical devices

Recall started
21 January 2020
Terminated
12 January 2023
Reason
The sterile pack inside the Complete Delivery System (CDS) did not go through a complete sterilization cycle, therefore the sterility of these components cannot be guaranteed.
Distribution
California only. No governmental, Canadian or Mexican consignees. TORRANCE MEMORIAL MEDICAL 3105 LOMITA BLVD TORRANCE CA 905055108 HOAG MEMORIAL HOSPITAL 1 HOAG DR NEWPORT BEACH CA 926634162 OWENS/MINORVMI TEM (65) 5125 ONTARIO MILLS PKWY ONTARIO CA 917645103
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1358-2020, Z-1357-2020

Terminated Class II Medical devices

Recall started
7 January 2020
Terminated
16 September 2020
Reason
There is potential for the sterile wrap to become loose while in transit and compromise the sterile barrier.
Distribution
Kentucky. No governmental, Canadian, or Mexican consignees.
Initiated
Voluntary: Firm initiated
Recall number
Z-0865-2020

Ongoing Class II Medical devices

Recall started
30 October 2019
Reason
The 10mm inner green inspiratory tubing is disconnected from the tee connector at the machine end of the circuit.
Distribution
USA - 2 consignees in PA only
Initiated
Voluntary: Firm initiated
Recall number
Z-0698-2020

Terminated Class II Medical devices

Recall started
18 September 2019
Terminated
6 May 2022
Reason
Incomplete seal and premature expiration of individually packaged Blood Glucose Test Strips.
Distribution
Nationwide distribution to CA, CT, GA, ID, IL, IN, KS, MA, MI, MN, MO, MS, NV, OH, OR, SC, TN, TX, UT, WA, WI, WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0513-2020

Ongoing Class II Medical devices

Recall started
6 September 2019
Reason
The recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2756-2019

Ongoing Class II Medical devices

Recall started
9 August 2019
Reason
Samples in the identified lot were found to have incomplete or open packaging seals.
Distribution
US nationwide distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2358-2019

Ongoing Class II Medical devices

Recall started
30 May 2019
Reason
The Anti-Fog Solution may not defog properly due to a degradation in the chemical properties.
Distribution
Nationwide distribution to AL, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, WY. Worldwide distribution to Canada and Malaysia.
Initiated
Voluntary: Firm initiated
Recall number
Z-1984-2019

Terminated Class II Medical devices

Recall started
28 May 2019
Terminated
30 January 2023
Reason
Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Distribution
US Nationwide distribution including in the states of CO, IL, KY, MO, NJ, OR, TN, TX, VA, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2043-2020, Z-2042-2020, Z-2044-2020

Terminated Class I Medical devices

Recall started
10 May 2019
Terminated
18 August 2021
Reason
This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.
Distribution
One distributor located in Illinois.
Initiated
Voluntary: Firm initiated
Recall number
Z-2067-2019

Terminated Class I Medical devices

Recall started
3 May 2019
Terminated
1 March 2021
Reason
The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.
Distribution
One distributor located in Illinois.
Initiated
Voluntary: Firm initiated
Recall number
Z-2066-2019

Terminated Class II Medical devices

Recall started
25 April 2019
Terminated
16 December 2020
Reason
Potential risk of brush detachment during cleaning. If brush detachment is not identified, there is the potential for the brush to be left within the scope.
Distribution
US Nationwide distribution in the states of AL, AR, CA, CO, DC, FL, GA, IA, IL, IN, KY, LA, MO, NC, NH, NJ, NM, NY, NV, OH, OK, PA, RI, TN, TX, UT, WV, and WY
Initiated
Voluntary: Firm initiated
Recall number
Z-2203-2019

Ongoing Class II Medical devices

Recall started
31 January 2019
Reason
Potential for the Semi-Rigid canister lid to fragment during use.
Distribution
Nationwide distribution to AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MT, NC, NH, NJ, NV, NY, OH, PA, RI, SC, TN, TX, VA, VT, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0174-2020

Terminated Class II Drugs

Recall started
17 September 2018
Terminated
4 September 2020
Reason
Microbial Contamination of Non-Sterile Product
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0247-2019

Terminated Class II Medical devices

Recall started
30 August 2018
Terminated
9 February 2023
Reason
Firm received sporadic reports of discoloration on certain lots of the product, which has been identified as common mold.
Distribution
United States, Taiwan, Chile
Initiated
Voluntary: Firm initiated
Recall number
Z-0088-2019

Terminated Class II Medical devices

Recall started
15 June 2018
Terminated
18 May 2020
Reason
There is potential risk of brush detachment during cleaning. If brush detachment is not identified, there is the potential for the brush to be left within the scope.
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, WY
Initiated
Voluntary: Firm initiated
Recall number
Z-3210-2018

Terminated Class II Medical devices

Recall started
24 April 2018
Terminated
4 August 2020
Reason
The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Saudi Arabia, Cayman Island, and United Arab Emirates
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1932-2018, Z-1936-2018, Z-1934-2018, Z-1935-2018, Z-1933-2018, Z-1931-2018

Terminated Class II Drugs

Recall started
15 February 2018
Terminated
29 April 2020
Reason
Subpotent Drug: product not meeting the iodine assay level requirements through the labeled expiry.
Distribution
Nationwide in the USA and Curacao
Initiated
Voluntary: Firm initiated
Recall number
D-0546-2018

Terminated Class II Medical devices

Recall started
1 February 2018
Terminated
5 October 2020
Reason
Packaged with an incorrect overwrap.
Distribution
US Nationwide Distribution to states of: CA, CT, DE, FL, GA, IN, LA, MA, MD, MN, MS, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, VT, WA and WV
Initiated
Voluntary: Firm initiated
Recall number
Z-2948-2018

Terminated Class II Medical devices

Recall started
12 January 2018
Terminated
22 July 2020
Reason
One of the cases of product might not have been sterilized.
Distribution
The product was distributed to UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-2378-2018

Terminated Class II Drugs

Recall started
21 September 2017
Terminated
28 August 2023
Reason
Subpotent Drug:The titratable iodine contained in the Povidone-Iodine prep pads is below label claim of 0.85%.
Distribution
Nationwide USA, Puerto Rico, China, Chile, Guam, Honduras, India, Israel, Malaysia, Mexico, Turkey
Initiated
Voluntary: Firm initiated
Recall numbers
D-0264-2018, D-0271-2018, D-0255-2018, D-0298-2018, D-0254-2018, D-0273-2018, D-0284-2018, D-0316-2018, D-0295-2018, D-0266-2018, D-0315-2018, D-0294-2018 and 56 more

Terminated Class II Medical devices

Recall started
18 July 2017
Terminated
24 April 2020
Reason
Product did not undergo correct sterilization procedures and may potentially be non-sterile.
Distribution
Worldwide Distribution - US Nationwide in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, WA and the countries of Canada, Saudi Arabia, and Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-0295-2018

Terminated Class II Medical devices

Recall started
12 April 2017
Terminated
5 August 2019
Reason
Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.
Distribution
CA, NM
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2976-2018, Z-2977-2018, Z-2979-2018, Z-2978-2018, Z-2980-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.