MEDLINE INDUSTRIES, LP - Northfield recalls

120 recalls recorded since 2018, most recently on 25 July 2025. 107 are listed as ongoing. Status and classification are as published by the FDA.

Recalls120
Ongoing (FDA)107
Terminated or completed13
Class I (FDA)12

Ongoing Class I Medical devices

Recall started
25 July 2025
Reason
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
Distribution
US distribution to: AL, CA, DE, OR, TN, TX
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2566-2025, Z-2567-2025

Ongoing Class II Medical devices

Recall started
23 July 2025
Reason
Kits contain SureStep Foley Tray Systems recalled by BD that include incorrect product information insert sheets. If a mislabeled tray is used, patient risk ranges from minor injury to injury that requires medical intervention, such as infection or local/systemic allergic reaction (anaphylaxis).
Distribution
Domestic distribution to FL and IL. No OUS distribution reported.
Initiated
Voluntary: Firm initiated
Recall number
Z-2479-2025

Ongoing Class I Medical devices

Recall started
14 July 2025
Reason
Specific lots of reprocessed electrophysiology catheters may contain small residual particulates.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-2460-2025

Ongoing Class II Medical devices

Recall started
8 July 2025
Reason
Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.
Distribution
US Nationwide distribution tin the states of CA, FL, IA, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2452-2025

Ongoing Class II Drugs

Recall started
7 July 2025
Reason
Subpotent Drug- isopropyl alcohol levels fall below the labeled concentration.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
D-0604-2025, D-0601-2025, D-0602-2025, D-0600-2025, D-0605-2025, D-0603-2025

Ongoing Class II Medical devices

Recall started
16 June 2025
Reason
SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-2199-2025

Ongoing Class II Medical devices

Recall started
12 June 2025
Reason
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms for the patient.
Distribution
Distributed domestically to California, Florida, Pennsylvania, and Texas. No international distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2155-2025, Z-2154-2025, Z-2156-2025, Z-2153-2025

Ongoing Class II Medical devices

Recall started
23 May 2025
Reason
Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.
Distribution
US Nationwide distribution in the states of IN, KY, CA, MI, TX, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2043-2025, Z-2044-2025, Z-2045-2025

Ongoing Class II Medical devices

Recall started
15 May 2025
Reason
There is a possible packaging defect impacting the outer Tyvec pouch that protects the sterile product.
Distribution
US Nationwide distribution in the state of VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1995-2025

Ongoing Class I Medical devices

Recall started
22 April 2025
Reason
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Distribution
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1809-2025, Z-1810-2025, Z-1811-2025

Ongoing Class II Medical devices

Recall started
9 April 2025
Reason
Three is the potential for the retractors to puncture through the sterile packaging.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1766-2025, Z-1765-2025

Ongoing Class II Medical devices

Recall started
24 March 2025
Reason
A single lot of the ImmunoCard STAT! Crypto/Giardia Test Kit was inadvertently stored in Medline warehouses at temperatures outside of the labeled storage requirements.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1719-2025

Ongoing Class II Medical devices

Recall started
19 March 2025
Reason
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1707-2025, Z-1709-2025, Z-1706-2025, Z-1705-2025, Z-1712-2025, Z-1713-2025, Z-1710-2025, Z-1711-2025, Z-1708-2025, Z-1714-2025

Ongoing Class II Medical devices

Recall started
5 March 2025
Reason
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Distribution
US Nationwide distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1548-2025, Z-1547-2025, Z-1546-2025, Z-1549-2025

Ongoing Class II Medical devices

Recall started
3 March 2025
Reason
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Panama, Virgin Islands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1605-2025, Z-1614-2025, Z-1618-2025, Z-1612-2025, Z-1599-2025, Z-1587-2025, Z-1611-2025, Z-1595-2025, Z-1601-2025, Z-1613-2025, Z-1610-2025, Z-1606-2025 and 22 more

Ongoing Class I Medical devices

Recall started
24 February 2025
Reason
Specific Medline Kits were manufactured using Intubation ORAL/NASAL Endotracheal Tubes which were subsequently recalled by Smiths Medical.
Distribution
California
Initiated
Voluntary: Firm initiated
Recall number
Z-1505-2025

Ongoing Class I Medical devices

Recall started
21 February 2025
Reason
Medical convenience kits containing Medtronic Aortic Root Cannula are being recalled due to the potential excess material in the male lures on the aortic cannula.
Distribution
US: CA, FL, MN, NY, OH, OR, PA, TX, WV
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1457-2025, Z-1459-2025, Z-1456-2025, Z-1458-2025

Ongoing Class II Medical devices

Recall started
19 February 2025
Reason
There is a potential for the sterility of the device to be compromised.
Distribution
Worldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: Bermuda, Canada, Panama, Singapore
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1531-2025, Z-1530-2025

Ongoing Class II Drugs

Recall started
17 February 2025
Reason
Subpotent Drug
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
D-0289-2025, D-0288-2025

Ongoing Class II Medical devices

Recall started
13 February 2025
Reason
Potential breach in pouch packaging which could lead to loss of sterility.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1364-2025, Z-1370-2025, Z-1369-2025, Z-1368-2025, Z-1366-2025, Z-1365-2025, Z-1363-2025, Z-1367-2025

Ongoing Class II Medical devices

Recall started
12 February 2025
Reason
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Distribution
Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1431-2025, Z-1428-2025, Z-1430-2025, Z-1429-2025

Ongoing Class II Medical devices

Recall started
11 February 2025
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1453-2025, Z-1439-2025, Z-1445-2025, Z-1441-2025, Z-1443-2025, Z-1435-2025, Z-1438-2025, Z-1447-2025, Z-1446-2025, Z-1452-2025, Z-1451-2025, Z-1449-2025 and 7 more

Ongoing Class II Medical devices

Recall started
11 February 2025
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1463-2025, Z-1464-2025, Z-1461-2025, Z-1465-2025, Z-1462-2025

Ongoing Class II Medical devices

Recall started
11 February 2025
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1394-2025, Z-1393-2025

Ongoing Class II Medical devices

Recall started
6 February 2025
Reason
Devices may have higher than expected amounts of bacterial endotoxin.
Distribution
US Nationwide distribution in the states of AR, AZ, CA, CO, IA, IL, KS, OR, TN, TX, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1460-2025

Ongoing Class II Medical devices

Recall started
9 January 2025
Reason
Kits contain recalled components due to reports of reduced skin barrier wear time. This refers to a shorter duration that a skin barrier can effectively adhere to and protect the skin before it needs to be replaced.
Distribution
US States: CA, NY
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1182-2025, Z-1183-2025

Ongoing Class II Medical devices

Recall started
3 January 2025
Reason
Medline is recalling medical convenience kits which were assembled with specific Coloplast Foley catheters and Prostatic catheters which were subsequently recalled due to a potential sterility issue detected during testing.
Distribution
US: CA
Initiated
Voluntary: Firm initiated
Recall number
Z-1173-2025

Ongoing Class I Medical devices

Recall started
31 December 2024
Reason
The catheter subassembly in the Integrated Arterial Catheter was manufactured with excess material on the catheter hub.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0992-2025, Z-0991-2025

Ongoing Class II Medical devices

Recall started
23 December 2024
Reason
Certain lots Sterile Infusion Catheters which were mislabeled as having a heparin coating were either distributed or manufactured into medical convenience kits and distributed.
Distribution
US: FL, KS, ND, PA, WA, WV, IL, WA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1049-2025, Z-1048-2025

Ongoing Class I Medical devices

Recall started
16 December 2024
Reason
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0765-2025, Z-0764-2025

Ongoing Class II Medical devices

Recall started
21 November 2024
Reason
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
Distribution
Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0778-2025, Z-0792-2025, Z-0793-2025, Z-0790-2025, Z-0791-2025, Z-0779-2025, Z-0777-2025, Z-0786-2025, Z-0789-2025, Z-0781-2025, Z-0787-2025, Z-0784-2025 and 6 more

Ongoing Class II Medical devices

Recall started
18 November 2024
Reason
Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.
Distribution
US and Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0697-2025, Z-0698-2025, Z-0696-2025, Z-0695-2025

Ongoing Class II Medical devices

Recall started
18 November 2024
Reason
Certain Medline kits were manufactured and distributed with Procedure Masks with SO SOFT Earloops, and SOFT TOUCH II Surgical Masks which were subsequently recalled for not meeting the bioburden testing requirements of EN 14683.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1042-2025, Z-1040-2025, Z-1041-2025

Ongoing Class II Medical devices

Recall started
12 November 2024
Reason
Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to increased risk of contamination and infection.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0758-2025, Z-0756-2025, Z-0757-2025, Z-0755-2025, Z-0760-2025, Z-0754-2025, Z-0759-2025

Ongoing Class II Medical devices

Recall started
6 November 2024
Reason
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Distribution
US Nationwide. Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0725-2025, Z-0722-2025, Z-0723-2025, Z-0719-2025, Z-0718-2025, Z-0721-2025, Z-0720-2025, Z-0726-2025, Z-0717-2025, Z-0724-2025

Ongoing Class II Medical devices

Recall started
25 October 2024
Reason
Identified lots of Olympus product may be missing sterile and manufacturing lot numbers and expiration dates on the sterile packaging. The information can be found on the outer box or zipper bag label. Medline is recalling its kits which utilize affected product as a component.
Distribution
US Nationwide distribution in the state of NY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0558-2025

Ongoing Class II Medical devices

Recall started
23 October 2024
Reason
Specimen container included in kits is labeled as sterile, but has been identified to be non-sterile. Long-term consequences may include unnecessary treatment and/or prolonged hospitalization.
Distribution
US Nationwide distribution in the states of AL, CA, FL, GA, IL, KY, MO, MS, MT, TN, TX, VA and WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0355-2025, Z-0359-2025, Z-0354-2025, Z-0356-2025, Z-0358-2025, Z-0360-2025, Z-0361-2025, Z-0357-2025

Ongoing Class II Medical devices

Recall started
21 October 2024
Reason
Tip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility.
Distribution
US Nationwide distributions in the states of AL, AR, CO, CT, FL, IA, IN, KS, LA, MA, MO, NC, NY, OK, OR, RI, TN, TX, VA, WA, WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0465-2025

Ongoing Class II Medical devices

Recall started
16 October 2024
Reason
The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position.
Distribution
US Nationwide. Bermuda, Canada, Panama, UAE, US Virgin Islands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0513-2025, Z-0510-2025, Z-0511-2025, Z-0509-2025, Z-0512-2025, Z-0515-2025, Z-0514-2025, Z-0508-2025

Ongoing Class II Medical devices

Recall started
15 October 2024
Reason
Some of the suture packages within the convenience kits contain the incorrect needle type/size and suture diameter/length.
Distribution
US Nationwide distribution in the states of AL, GA, MA, NJ, and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0533-2025, Z-0532-2025, Z-0534-2025

Ongoing Class II Medical devices

Recall started
9 October 2024
Reason
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0632-2025, Z-0623-2025, Z-0624-2025, Z-0639-2025, Z-0645-2025, Z-0636-2025, Z-0625-2025, Z-0640-2025, Z-0637-2025, Z-0634-2025, Z-0643-2025, Z-0644-2025 and 16 more

Ongoing Class II Medical devices

Recall started
4 October 2024
Reason
Products within a Medline pack may be impacted by a packaging error in which an incomplete seal has formed on the packaging thereby creating a packaging seal gap.
Distribution
US Nationwide distribution in the state of TN.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0349-2025, Z-0350-2025

Ongoing Class II Medical devices

Recall started
4 September 2024
Reason
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0175-2025, Z-0165-2025, Z-0164-2025, Z-0167-2025, Z-0159-2025, Z-0156-2025, Z-0151-2025, Z-0180-2025, Z-0172-2025, Z-0179-2025, Z-0178-2025, Z-0153-2025 and 21 more

Ongoing Class II Medical devices

Recall started
3 September 2024
Reason
Product has been identified as having holes in the packaging.
Distribution
Worldwide distribution - US Nationwide and the countries of India and Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-0133-2025

Ongoing Class II Medical devices

Recall started
30 August 2024
Reason
It has been determined that certain Blood Pressure Monitors containing a faulty microchip have failed to power on or have shown display screen issues after the initial few uses, as well as overheating at times making the unit warm to the touch.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama and Jamaica.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0034-2025, Z-0033-2025, Z-0032-2025

Ongoing Class II Medical devices

Recall started
28 August 2024
Reason
Medline Industries, LP. has identified certain kits containing the recalled Medtronic lots of the DLP Vessel Cannulae. During manufacturing of the three listed lot numbers, product for model 30001 was incorrectly labeled as model 30000.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0123-2025, Z-0122-2025, Z-0121-2025, Z-0119-2025, Z-0120-2025, Z-0124-2025

Ongoing Class II Medical devices

Recall started
28 August 2024
Reason
CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.
Distribution
US Nationwide distribution in the state of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0053-2025

Ongoing Class II Medical devices

Recall started
27 June 2024
Reason
XXX
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3120-2024, Z-3118-2024, Z-3106-2024, Z-3119-2024, Z-3111-2024, Z-3110-2024, Z-3114-2024, Z-3109-2024, Z-3115-2024, Z-3107-2024, Z-3105-2024, Z-3121-2024 and 5 more

Ongoing Class II Medical devices

Recall started
18 June 2024
Reason
A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
Distribution
US: TX, IL, CA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3069-2024, Z-3072-2024, Z-3070-2024, Z-3071-2024

Ongoing Class II Medical devices

Recall started
7 May 2024
Reason
Potential for an incomplete seal on the packaging tray.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3093-2024, Z-3092-2024, Z-3094-2024, Z-3095-2024

Ongoing Class II Medical devices

Recall started
30 April 2024
Reason
There is the potential of the tip protector to fall off 4.5" sterile, curved, Iris Scissors
Distribution
Worldwide distribution - US Nationwide and the countries of Singapore, Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2024-2024

Ongoing Class II Medical devices

Recall started
22 April 2024
Reason
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1938-2024, Z-1937-2024

Ongoing Class II Medical devices

Recall started
17 April 2024
Reason
These syringes were identified to be affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-1936-2024

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2754-2024, Z-2739-2024, Z-2752-2024, Z-2743-2024, Z-2741-2024, Z-2744-2024, Z-2747-2024, Z-2759-2024, Z-2764-2024, Z-2765-2024, Z-2755-2024, Z-2756-2024 and 15 more

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2775-2024, Z-2774-2024, Z-2776-2024, Z-2780-2024, Z-2779-2024, Z-2778-2024, Z-2777-2024

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2624-2024, Z-2630-2024, Z-2632-2024, Z-2635-2024, Z-2629-2024, Z-2625-2024, Z-2627-2024, Z-2626-2024, Z-2623-2024, Z-2628-2024, Z-2631-2024, Z-2634-2024 and 1 more

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3056-2024, Z-2990-2024, Z-3050-2024, Z-3038-2024, Z-3023-2024, Z-3047-2024, Z-3030-2024, Z-3003-2024, Z-3013-2024, Z-2984-2024, Z-3010-2024, Z-3037-2024 and 69 more

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2768-2024, Z-2773-2024, Z-2770-2024, Z-2767-2024, Z-2766-2024, Z-2769-2024, Z-2771-2024, Z-2772-2024

Ongoing Class I Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide Distribution: US (nationwide) and OUS countries of: AE, AI, BB, CA, CL, CY, DE, GT, HN, JP, MX, MY, PA, PE, PR, SA, SG, TC, TT, VI, PR, TT, MX, BM, ES, NI, DO, GU, CR, AT, CD, DE, DF, W, and ON.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2443-2024, Z-2444-2024, Z-2442-2024

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2516-2024, Z-2515-2024

Ongoing Class II Medical devices

Recall started
22 March 2024
Reason
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
Distribution
Worldwide - US Nationwide and the countries of Panama, Mexico, Canada, Qatar, El Salvador, Oman, Colombia, Singapore.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2694-2024, Z-2701-2024, Z-2695-2024, Z-2704-2024, Z-2703-2024, Z-2699-2024, Z-2708-2024, Z-2710-2024, Z-2697-2024, Z-2709-2024, Z-2702-2024, Z-2696-2024 and 9 more

Ongoing Class II Medical devices

Recall started
14 March 2024
Reason
Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.
Distribution
US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2228-2024, Z-2230-2024, Z-2226-2024, Z-2232-2024, Z-2231-2024, Z-2224-2024, Z-2229-2024, Z-2227-2024, Z-2225-2024

Ongoing Class II Medical devices

Recall started
12 March 2024
Reason
Products have a lack of sterility assurance.
Distribution
US Nationwide distribution in the states of AL, FL, KY, MS, TX, WA, WI, NC, VA, CA, SD, TN, GA, PA, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1807-2024, Z-1806-2024

Ongoing Class II Medical devices

Recall started
5 March 2024
Reason
Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory compliance requirement (ASTM F3186-17, Standard Specification for Adult Portable Bed Rails and Related Products and 16 CFR Part 1270, Safety Standard for Adult Portable Bed Rails) and failed to meet all mandatory requirements outlined in these standards.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1487-2024, Z-1488-2024

Ongoing Class II Medical devices

Recall started
1 March 2024
Reason
Trumpet Needle Guide ring can detach when excessive pressure is applied.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1918-2024, Z-1917-2024

Ongoing Class I Medical devices

Recall started
26 February 2024
Reason
Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.
Distribution
Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall number
Z-1876-2024

Ongoing Class II Medical devices

Recall started
26 February 2024
Reason
The firm provided a set of specifications for the device, but some units of the device were manufactured non-compliant with the specifications. However, the firms Receiving Dept. accepted these units into inventory
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1459-2024

Ongoing Class II Medical devices

Recall started
23 February 2024
Reason
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama, Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1444-2024, Z-1451-2024, Z-1429-2024, Z-1438-2024, Z-1453-2024, Z-1452-2024, Z-1446-2024, Z-1434-2024, Z-1440-2024, Z-1455-2024, Z-1443-2024, Z-1435-2024 and 17 more

Ongoing Class II Medical devices

Recall started
23 January 2024
Reason
Product failed sterility testing.
Distribution
US Nationwide distribution in the states of AL, CA, FL, IN, MI, MO, NJ, NY, PA, RI, SC, TN, TX. and VA. There was no foreign/government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1295-2024

Ongoing Class II Drugs

Recall started
10 January 2024
Reason
Labeling: Not Elsewhere Classified; Product labeling contains the claim of 'Paraben Free' while the product does in fact contain parabens as part of the formulation/ ingredient list.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0315-2024

Ongoing Class II Medical devices

Recall started
4 January 2024
Reason
Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).
Distribution
US, Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1157-2024

Ongoing Class II Medical devices

Recall started
21 December 2023
Reason
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Various Adult Blood and Blood Culture Kits. This recall is due to kits built using a Bactec component lot and expiration date that were not reflected on the kit insert. There is a risk of using an expired component if the user does not check the Bactec inside the kit for its expiration date.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0967-2024, Z-0968-2024, Z-0965-2024, Z-0964-2024, Z-0963-2024, Z-0969-2024, Z-0966-2024

Ongoing Class II Medical devices

Recall started
20 December 2023
Reason
There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.
Distribution
US and territories, S. Korea, Panama, and Dubai
Initiated
Voluntary: Firm initiated
Recall number
Z-1180-2024

Ongoing Class II Medical devices

Recall started
29 November 2023
Reason
Surgical gowns were manufactured with the wrong sleeve.
Distribution
Distribution was made to MD, OH, TN, TX, and WI. There was no government/military/foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0711-2024

Ongoing Class I Medical devices

Recall started
15 November 2023
Reason
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
Distribution
Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1090-2024, Z-1124-2024, Z-1118-2024, Z-1134-2024, Z-1123-2024, Z-1102-2024, Z-1122-2024, Z-1116-2024, Z-1114-2024, Z-1104-2024, Z-1096-2024, Z-1119-2024 and 37 more

Ongoing Class II Medical devices

Recall started
3 November 2023
Reason
The product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.
Distribution
US Nationwide distribution in the states of CT, IA, LA, MA, PA, CA, NY, FL.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0525-2024, Z-0531-2024, Z-0530-2024, Z-0528-2024, Z-0527-2024, Z-0526-2024, Z-0529-2024

Ongoing Class II Medical devices

Recall started
1 November 2023
Reason
This recall is being issued due to items within the kit have expired prior to the overall kit expiration date listed on the insert.
Distribution
US Nationwide distribution in the state of California.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0310-2024, Z-0311-2024, Z-0314-2024, Z-0313-2024, Z-0312-2024

Ongoing Class II Medical devices

Recall started
19 October 2023
Reason
Undeclared latex
Distribution
US and Panama
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0535-2024, Z-0533-2024, Z-0534-2024

Ongoing Class I Medical devices

Recall started
16 October 2023
Reason
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Distribution
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0561-2024, Z-0562-2024, Z-0563-2024

Terminated Class II Drugs

Recall started
4 October 2023
Terminated
30 September 2024
Reason
CGMP deviations: the product was shipped from the Manufacturer to a Medline warehouse and released to stock while it was still under investigation for low assay results on the active ingredient miconazole nitrate.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0091-2024

Ongoing Class I Medical devices

Recall started
4 August 2023
Reason
Product dispositioned for scrap for sterility failure investigation that was inadvertently shipped to customers.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2514-2023

Ongoing Class II Medical devices

Recall started
3 August 2023
Reason
4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore
Initiated
Voluntary: Firm initiated
Recall number
Z-2595-2023

Ongoing Class II Medical devices

Recall started
3 August 2023
Reason
Various reprocessed Covidien LigaSure and Ethicon Harmonic devices may have an incomplete seal on the packaging tray.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0120-2024, Z-0118-2024, Z-0119-2024, Z-0117-2024

Ongoing Class II Medical devices

Recall started
3 July 2023
Reason
The prep solution included in the Total Hip Kit expires prior to the expiration date of the Kit.
Distribution
US Nationwide distribution in the state of Illinois.
Initiated
Voluntary: Firm initiated
Recall number
Z-2366-2023

Ongoing Class II Medical devices

Recall started
23 May 2023
Reason
The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2649-2023

Ongoing Class II Medical devices

Recall started
18 May 2023
Reason
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Distribution
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2176-2023, Z-2169-2023, Z-2163-2023, Z-2195-2023, Z-2205-2023, Z-2161-2023, Z-2203-2023, Z-2159-2023, Z-2202-2023, Z-2194-2023, Z-2204-2023, Z-2192-2023 and 40 more

Ongoing Class II Medical devices

Recall started
18 May 2023
Reason
Affected product lots were mislabeled as "STERILE PVP SLN" when they should have been titled "NON-STERILE PVP SOLUTION, STERILE PACKAGING".
Distribution
Worldwide distribution - US Nationwide and the country of Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2134-2023, Z-2133-2023, Z-2131-2023, Z-2132-2023, Z-2135-2023, Z-2137-2023, Z-2136-2023

Ongoing Class II Medical devices

Recall started
15 May 2023
Reason
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2305-2023, Z-2329-2023, Z-2321-2023, Z-2336-2023, Z-2306-2023, Z-2334-2023, Z-2338-2023, Z-2298-2023, Z-2328-2023, Z-2333-2023, Z-2309-2023, Z-2300-2023 and 38 more

Ongoing Class II Medical devices

Recall started
12 May 2023
Reason
The scissors were manufactured without a tip protector resulting in the scissors breaking through the semi-rigid plastic tray compromising sterility.
Distribution
US Nationwide distribution. There was government distribution but no military or foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2206-2023

Ongoing Class II Medical devices

Recall started
3 May 2023
Reason
Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
Distribution
Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1982-2023

Ongoing Class II Medical devices

Recall started
21 April 2023
Reason
The kits were damaged by water.
Distribution
Distribution was made to California. There was no foreign/government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1646-2023

Ongoing Class II Medical devices

Recall started
28 March 2023
Reason
The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date.
Distribution
Worldwide distribution - US Nationwide and the country of Bermuda.
Initiated
Voluntary: Firm initiated
Recall number
Z-2227-2023

Ongoing Class II Medical devices

Recall started
2 March 2023
Reason
The CirClamp subassembly found in the kit was the incorrect size.
Distribution
US Nationwide distribution in the states of CA and TX. There was no foreign/government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1407-2023

Ongoing Class II Medical devices

Recall started
27 December 2022
Reason
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
Distribution
US Nationwide distribution in the states of AR, CO, and GA. There was no foreign/military/government distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1163-2023, Z-1161-2023, Z-1162-2023

Ongoing Class II Medical devices

Recall started
23 December 2022
Reason
The kit contained an expired component.
Distribution
US Nationwide distribution in the states of NY, IN, FL, NJ, IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-1015-2023

Ongoing Class II Medical devices

Recall started
22 December 2022
Reason
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1047-2023, Z-1045-2023, Z-1046-2023

Ongoing Class II Medical devices

Recall started
7 December 2022
Reason
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1125-2023, Z-1117-2023, Z-1114-2023, Z-1111-2023, Z-1123-2023, Z-1113-2023, Z-1112-2023, Z-1109-2023, Z-1124-2023, Z-1118-2023, Z-1119-2023, Z-1115-2023 and 8 more

Ongoing Class II Medical devices

Recall started
18 November 2022
Reason
Potential of the blade puncturing the outer foil layer causing a breach in the sterile packaging.
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, MD, MI, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0930-2023

Ongoing Class II Medical devices

Recall started
17 November 2022
Reason
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1077-2023, Z-1064-2023, Z-1087-2023, Z-1072-2023, Z-1092-2023, Z-1067-2023, Z-1094-2023, Z-1074-2023, Z-1062-2023, Z-1071-2023, Z-1065-2023, Z-1095-2023 and 26 more

Ongoing Class II Medical devices

Recall started
17 November 2022
Reason
Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.
Distribution
US Nationwide distribution in the states of AR, MD, WA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1032-2023, Z-1033-2023, Z-1029-2023, Z-1035-2023, Z-1034-2023, Z-1031-2023, Z-1030-2023

Ongoing Class II Medical devices

Recall started
30 September 2022
Reason
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
Distribution
US Nationwide distribution in the states of CA, MN, NC, NY, OK, PA, TN, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0167-2023, Z-0164-2023, Z-0166-2023, Z-0165-2023, Z-0157-2023, Z-0160-2023, Z-0162-2023, Z-0156-2023, Z-0155-2023, Z-0161-2023, Z-0163-2023, Z-0158-2023 and 1 more

Ongoing Class II Medical devices

Recall started
9 September 2022
Reason
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match the date listed on the kit.
Distribution
US Nationwide distribution in the state of California.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0120-2023, Z-0121-2023, Z-0124-2023, Z-0122-2023, Z-0123-2023

Terminated Class II Medical devices

Recall started
7 September 2022
Terminated
4 March 2026
Reason
One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
Distribution
US Nationwide distribution in the states of CA and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0106-2023

Terminated Class II Medical devices

Recall started
1 September 2022
Terminated
26 January 2026
Reason
Impacted Venclose Procedure Packs contain a 12cm introducer/sheath component, either component #128626 - SET INTRO 7FX12CM FITS 0.018 or component #137339 - SET INTRO 6.5FX12CM FITS 0.018, instead of a 7cm introducer/sheath, component #137340 - SET INTRO 6.5FX7CM FITS 0.018
Distribution
CA
Initiated
Voluntary: Firm initiated
Recall number
Z-0088-2023

Ongoing Class II Medical devices

Recall started
3 August 2022
Reason
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Pa
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1857-2022, Z-1856-2022, Z-1855-2022

Ongoing Class II Medical devices

Recall started
29 June 2022
Reason
stability failure
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1493-2022

Ongoing Class II Medical devices

Recall started
11 May 2022
Reason
Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Singapore, Panama, and Hongkong.
Initiated
Voluntary: Firm initiated
Recall number
Z-1287-2022

Ongoing Class II Medical devices

Recall started
21 April 2022
Reason
Sporadic reports were received of discoloration on the foam sponge on certain lots of the product, which has been identified as microbial contamination.
Distribution
US and Panama
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1218-2022, Z-1219-2022, Z-1220-2022, Z-1221-2022, Z-1217-2022

Terminated Class II Medical devices

Recall started
9 March 2022
Terminated
26 May 2026
Reason
The mask cuff may disconnect from the device's breathing tube.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0941-2022

Terminated Class II Medical devices

Recall started
8 March 2022
Terminated
27 February 2023
Reason
The product was stored incorrectly due to improper storage controls which may lead to delayed results.
Distribution
Distribution was made to AL, FL, NE OH, TX, WA, and WI. There was no foreign/military/government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0933-2022

Terminated Class II Medical devices

Recall started
8 March 2022
Terminated
20 March 2023
Reason
Product was improperly store which may cause a delay in results.
Distribution
US distribution only - AL, CA, FL, GA, IL, IN, LA, MO, MS, NC, NE, NY, OH, PA, VA, and WA
Initiated
Voluntary: Firm initiated
Recall number
Z-0932-2022

Terminated Class II Medical devices

Recall started
4 March 2022
Terminated
20 March 2023
Reason
Sterility failure of the exterior of the syringe which may impact the device's ability to be laid on a sterile field.
Distribution
US Nationwide Distribution was made to KY, OK, and TX. There was no foreign/government/military distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0898-2022

Terminated Class II Medical devices

Recall started
25 January 2022
Terminated
9 January 2023
Reason
The product was incorrectly shipped at room temperature rather than refrigerated.
Distribution
US Distribution was made to IL, FL, LA, MS, NC, OH, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0678-2022

Ongoing Class II Medical devices

Recall started
29 December 2021
Reason
Micro-Kill Bleach Wipes contain out of specification (low) levels of Sodium Hypochlorite.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama and Bermuda.
Initiated
Voluntary: Firm initiated
Recall number
Z-0696-2022

Ongoing Class II Medical devices

Recall started
15 December 2021
Reason
The product contains natural rubber latex but the label states that it does not.
Distribution
US and Bahamas, Bermuda, Panama, Lebanon, Barbados, Honduras, Anguilla
Initiated
Voluntary: Firm initiated
Recall number
Z-0633-2022

Terminated Class II Medical devices

Recall started
1 December 2021
Terminated
9 January 2023
Reason
Product stored incorrectly in temperature controlled setting instead of refrigeration.
Distribution
US Nationwide distribution in the state of California.
Initiated
Voluntary: Firm initiated
Recall number
Z-0554-2022

Terminated Class II Medical devices

Recall started
1 December 2021
Terminated
13 February 2023
Reason
Product stored incorrectly in temperature controlled setting instead of refrigeration.
Distribution
US Nationwide distribution in the states of California and Massachusetts.
Initiated
Voluntary: Firm initiated
Recall number
Z-0552-2022

Terminated Class II Medical devices

Recall started
1 December 2021
Terminated
13 February 2023
Reason
Product stored incorrectly in temperature controlled setting instead of refrigeration.
Distribution
US Nationwide distribution in the states of Colorado, Florida, New York, North Carolina, and Texas.
Initiated
Voluntary: Firm initiated
Recall number
Z-0656-2022

Terminated Class II Medical devices

Recall started
30 November 2021
Terminated
6 January 2023
Reason
Affected product was shipped from the warehouse at room temperature instead of the required refrigerated environment.
Distribution
US Nationwide distribution in the states of IL, OH, MS, and PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0404-2022

Terminated Class II Medical devices

Recall started
22 October 2021
Terminated
4 January 2023
Reason
Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.
Distribution
US Distribution to North Carolina.
Initiated
Voluntary: Firm initiated
Recall number
Z-1140-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.