Medicrea International recalls

9 recalls recorded since 2018, most recently on 29 October 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)4
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
29 October 2025
Reason
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
Distribution
US Nationwide distribution in the states of TN, CA, MN, NC, TX, FL, OH, RI, VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1039-2026, Z-1038-2026

Ongoing Class II Medical devices

Recall started
3 December 2024
Reason
Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland
Initiated
Voluntary: Firm initiated
Recall number
Z-1118-2025

Ongoing Class II Medical devices

Recall started
12 July 2023
Reason
There is a potential for a product packaging non-conformity issue which presents as a pinhole in either the inner or outer pouch.
Distribution
US nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2461-2023, Z-2460-2023, Z-2456-2023, Z-2453-2023, Z-2459-2023, Z-2462-2023, Z-2455-2023, Z-2452-2023, Z-2457-2023, Z-2458-2023, Z-2454-2023

Terminated Class II Medical devices

Recall started
8 October 2020
Terminated
9 May 2024
Reason
Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,
Distribution
CA Foreign:France, Belgium
Initiated
Voluntary: Firm initiated
Recall number
Z-0463-2021

Terminated Class II Medical devices

Recall started
28 January 2020
Terminated
17 March 2023
Reason
An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1472-2020, Z-1473-2020

Terminated Class II Medical devices

Recall started
11 November 2019
Terminated
14 October 2020
Reason
Incompatible universal counter torque (A22020500) with the nut driver T30 (A22020220)
Distribution
CA, FL
Initiated
Voluntary: Firm initiated
Recall number
Z-1234-2020

Terminated Class II Medical devices

Recall started
2 October 2019
Terminated
8 February 2021
Reason
Inner draw rod, which threads into the cage may potentially be separated from the proximal knob of the inserter
Distribution
OH, VA
Initiated
Voluntary: Firm initiated
Recall number
Z-1156-2020

Ongoing Class II Medical devices

Recall started
16 September 2019
Reason
Conical Cannulated Screws may not accept 1.6mm K-Wire, it cannot enter the minimum tolerance axis threaded
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0827-2020, Z-0828-2020

Terminated Class II Medical devices

Recall started
27 June 2019
Terminated
19 November 2020
Reason
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
Distribution
US: NY, CO, OH, AZ, FL, TX, PA OUS: France, Belgium, South Africa, China, Spain, Portugal
Initiated
Voluntary: Firm initiated
Recall number
Z-0753-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.