Recall Record

Milan, N/A, Italy

Medicalplastic S.R.L. recalls

1 recall recorded since 2018, most recently on 17 June 2021. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
17 June 2021
Reason
Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Distribution
Memphis, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-0245-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.