Bedford, MA, United States

Medica Corporation recalls

4 recalls recorded since 2018, most recently on 31 December 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)1
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
31 December 2025
Reason
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
Distribution
International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
Initiated
Voluntary: Firm initiated
Recall number
Z-1370-2026

Terminated Class II Medical devices

Recall started
6 February 2019
Terminated
2 October 2020
Reason
Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
Distribution
Worldwide distribution - US Nationwide and countries of IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE GUATEMALA, KENYA, IRAN, ESTONIA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1294-2019

Terminated Class II Medical devices

Recall started
21 November 2017
Terminated
20 August 2020
Reason
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.
Distribution
Worldwide and US Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1149-2018

Terminated Class II Medical devices

Recall started
11 April 2017
Terminated
20 August 2019
Reason
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
Distribution
USA: MI FL CA TX India, Russia, Puerto Rico, N. Mariana Island/Palau.
Initiated
Voluntary: Firm initiated
Recall number
Z-1267-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.