Maquet Cardiovascular Us Sales, Llc recalls

15 recalls recorded since 2018, most recently on 20 November 2020. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls15
Ongoing (FDA)0
Terminated or completed15
Class I (FDA)2

Terminated Class II Medical devices

Recall started
20 November 2020
Terminated
7 November 2024
Reason
The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product's sterility barrier.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Austria, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Ko
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0632-2021, Z-0633-2021

Terminated Class II Medical devices

Recall started
30 September 2020
Terminated
25 May 2023
Reason
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, IA, MD, NC, PA, TX, WA, WI and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Italy, Jordan, Malaysia,
Initiated
Voluntary: Firm initiated
Recall number
Z-0443-2021

Terminated Class II Medical devices

Recall started
24 September 2020
Terminated
28 February 2023
Reason
Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
Distribution
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, K
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0132-2021, Z-0133-2021

Terminated Class III Medical devices

Recall started
7 August 2020
Terminated
14 January 2021
Reason
Instruction For Use (IFU) references incorrect serial number of 90437000. It should state This document applies to ROTAFLOW Consoles as of serial number 94175452 affected by the implementation of the new IEC standard 60601 1 2, 4th edition
Distribution
Worldwide distribution - US Nationwide distribution including in the states of ME, NY, NV, OH, TX, VA, WA and the countries of China, Colombia, India, Italy, Mexico, Singapore, South Korea, Spain, Taiwan, Thailand, Turkey, United Arabic Emirates, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2957-2020

Terminated Class II Medical devices

Recall started
18 June 2020
Terminated
30 December 2022
Reason
Removal for a specific lot, lot number 70128207, of Holder HKHZ 19 for use with QUADROX iD Pediatric Oxygenator. The Holder may not always snap and securely hold the oxygenator as intended.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of OH, PA, SC, CT and the countries of Germany, Spain, France, United Kingdom, Italy, United Arab Emirates, Mexico, Canada, Croatia, Georgia, Poland, Slovakia, Thailand, and Vietnam.
Initiated
Voluntary: Firm initiated
Recall number
Z-2780-2020

Terminated Class II Medical devices

Recall started
20 February 2020
Terminated
18 January 2022
Reason
A lot of Avalon Elite Vascular Access Kits (VAK) are being recalled due to a material mix-up with the needle portion of the kit; 81 of 255 kits were assembled with an incorrect Arterial Access Needle instead of a Multi-Purpose Procedure Needle.
Distribution
Domestic distribution only to the following states: AL AZ CA CT FL GA IL MA MI NC NH NM NY OH OK TN TX VA
Initiated
Voluntary: Firm initiated
Recall number
Z-1603-2020

Terminated Class II Medical devices

Recall started
10 February 2020
Terminated
13 September 2023
Reason
Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water
Distribution
Nationwide Foreign: Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Iraq, Iran, Ireland, Israel, Italy, Japan, Kenya, Lithuania, Netherlands, New Zealand, Norway, Poland, Qatar, Saudi Arabia, Singapore, Slovenia, Sou
Initiated
Voluntary: Firm initiated
Recall number
Z-1475-2020

Terminated Class III Medical devices

Recall started
16 January 2020
Terminated
23 October 2023
Reason
Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
Distribution
U.S. Distribution: CA, CO, IA, MN, NY, SC, VA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1390-2020, Z-1391-2020

Terminated Class II Medical devices

Recall started
3 December 2019
Terminated
18 September 2020
Reason
A loose ODU plug / Coaxial cable connection may result in fluctuating flow values on the Console.
Distribution
The products were distributed to the following US states: ME, NV, NY, OH, PA, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0678-2020

Terminated Class II Medical devices

Recall started
11 July 2019
Terminated
11 December 2020
Reason
The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.
Distribution
The products were distributed to the following US states: AR, CA, FL, GA, IA, IL, IN, KY, NC, NE, NM, NV, NY, OH, OR, TN, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2149-2019

Terminated Class II Medical devices

Recall started
28 May 2019
Terminated
10 July 2023
Reason
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
Distribution
Worldwide distribution. US Nationwide, ALBANIA GHANA PORTUGAL ARGENTINA HONG KONG ROMANIA AUSTRIA HUNGARY RUSSIA AUSTRALIA INDIA SAUDI ARABIA AZERBAIJAN IRAN SINGAPORE BAHAMAS IRELAND SLOVAKIA BANGLADESH IRAQ SLOVENIA BELGIUM ISRAEL SOUTH AFRICA BOLIVIA ITALY SOUTH KOREA BRAZIL JAPAN SPAIN BULG
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1864-2019, Z-1860-2019, Z-1857-2019, Z-1856-2019, Z-1858-2019, Z-1855-2019, Z-1859-2019, Z-1852-2019, Z-1853-2019, Z-1861-2019, Z-1863-2019, Z-1854-2019 and 1 more

Terminated Class II Medical devices

Recall started
20 May 2019
Terminated
15 December 2023
Reason
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Distribution
Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1777-2019, Z-1778-2019

Terminated Class II Medical devices

Recall started
6 May 2019
Terminated
6 July 2022
Reason
User may suffer an electric shock when coming in contact with the exposed electronic components
Distribution
Worldwide - US Nationwide Distribution Foreign: ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BOLIVIA BOTSWANA BOSNIA AND HERZEGOVINA BRAZIL BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK ECUADOR EGYPT FINLAND FRANCE
Initiated
Voluntary: Firm initiated
Recall number
Z-1734-2019

Terminated Class I Medical devices

Recall started
8 February 2019
Terminated
14 May 2020
Reason
Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.
Distribution
US Nationwide Distribution in the states of California, Florida, Illinois, Kentucky, Michigan, and New York.
Initiated
Voluntary: Firm initiated
Recall number
Z-0986-2019

Terminated Class I Medical devices

Recall started
25 October 2018
Terminated
5 June 2020
Reason
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakist
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0436-2019, Z-0437-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.