Mako Surgical Corporation recalls

9 recalls recorded since 2018, most recently on 8 August 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)0
Terminated or completed9
Class I (FDA)0

Terminated Class II Medical devices

Recall started
8 August 2022
Terminated
22 May 2026
Reason
Certain units have the potential to exhibit infant failures whereby units either arrive non-functional, have the potential to become non-functional, or flash an error light while in use.
Distribution
US Distribution was made to IL and AZ.
Initiated
Voluntary: Firm initiated
Recall number
Z-1735-2022

Completed Class II Medical devices

Recall started
23 October 2020
Reason
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK
Initiated
Voluntary: Firm initiated
Recall number
Z-0472-2021

Terminated Class II Medical devices

Recall started
9 July 2020
Terminated
16 June 2023
Reason
Characterization issue associated with the product. This issue may cause a discrepancy during bone preparation in certain Mako Total Hip procedures.
Distribution
TX, MN, CA, RI, FL, NY, AZ, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-2745-2020

Terminated Class II Medical devices

Recall started
25 April 2019
Terminated
22 April 2021
Reason
The bearings of the saw attachments were ungreased.
Distribution
AR, CA, FL, KS, MA, MO, NC, NJ, NY, OH, OR, TN, and TX. Australia, China and Netherlands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1823-2019, Z-1824-2019

Terminated Class II Medical devices

Recall started
24 July 2018
Terminated
11 October 2022
Reason
Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera.
Distribution
All devices were distributed in the US in the following states: AK, AL, AR, AZ, CA, CO, CT, FL, IA, ID, KS, KY, LA, MA, MD, MI, MN, MO, NE, NC, NH, NJ, NY, NV, OH, PA, RI, SC, SD, TN. TX, UT, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2924-2018, Z-2925-2018

Terminated Class II Medical devices

Recall started
10 May 2018
Terminated
17 January 2021
Reason
Out of tolerance.
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY and the countries of Australia, Hong Kong, India, Italy, Korea, Midrand Allandal
Initiated
Voluntary: Firm initiated
Recall number
Z-2794-2018

Terminated Class II Medical devices

Recall started
19 January 2018
Terminated
16 September 2020
Reason
Possible hinge pin disassociation from the Mako Onlay Insert Extractor.
Distribution
Distributed domestically to AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, WI. Distributed internationally to Australia, China, Italy, Netherlands, Singapore, South Africa, Taiwan, Turkey, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2051-2018

Terminated Class II Medical devices

Recall started
27 December 2017
Terminated
14 December 2020
Reason
Communication-connection error.
Distribution
Worldwide Distribution - US Nationwide and the countries of Australia, New Zealand, Taiwan, China, Hong Kong, Greece, United Kingdom, England, Netherlands, Luxembourg, Germany, Ireland, Switzerland, Italy, India, Japan, South Korea, Vietnam, South Africa, and Turkey,
Initiated
Voluntary: Firm initiated
Recall number
Z-3182-2018

Terminated Class II Medical devices

Recall started
1 December 2017
Terminated
14 December 2020
Reason
Incorrect product and/or label. Sizing is labeled incorrectly.
Distribution
AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OK, OR, PA, RI, SC, TX, UT, VA, WA, WY, and Hawaii Australia, Italy, Netherlands, Singapore, South Korea, United Kingdom, and Vietnam
Initiated
Voluntary: Firm initiated
Recall number
Z-1644-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.