Recall Record

North Billerica, MA, United States

Magellan Diagnostics, Inc. recalls

7 recalls recorded since 2018, most recently on 4 September 2026. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)4
Terminated or completed3
Class I (FDA)1

Ongoing Class III Medical devices

Recall started
4 September 2026
Reason
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
Distribution
US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0162-2027

Ongoing Class II Medical devices

Recall started
13 March 2026
Reason
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, Sou
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1894-2026, Z-1893-2026, Z-1895-2026

Ongoing Class II Medical devices

Recall started
23 February 2024
Reason
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0584-2026

Ongoing Class II Medical devices

Recall started
19 September 2023
Reason
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0219-2024

Terminated Class I Medical devices

Recall started
7 May 2021
Terminated
24 October 2023
Reason
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1954-2021, Z-1953-2021, Z-1952-2021

Terminated Class II Medical devices

Recall started
8 August 2018
Terminated
12 May 2020
Reason
Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.
Distribution
Worldwide distribution: US (Nationwide) and countries of: Spain and South Africa.
Initiated
Voluntary: Firm initiated
Recall number
Z-0285-2019

Terminated Class II Medical devices

Recall started
9 January 2018
Terminated
11 December 2020
Reason
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1295-2018, Z-1296-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.