Recall Record

Green Bay, WI, United States

LITTLE RAPIDS CORPORATION recalls

1 recall recorded since 2018, most recently on 27 December 2017. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
27 December 2017
Terminated
30 July 2018
Reason
The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.
Distribution
TX, CT, TN, MO, WI, FL, SD, MS, CA, MS
Initiated
Voluntary: Firm initiated
Recall number
Z-0578-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.