Recall Record

Saint Louis, MO, United States

LIEBEL-FLARSHEIM COMPANY LLC recalls

3 recalls recorded since 2018, most recently on 31 July 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)1

Ongoing Class I Drugs

Recall started
31 July 2026
Reason
Presence of Particulate Matter: comprising polyethylene and other plastic materials, stainless steel and glass.
Distribution
Distributed Nationwide in the US and Mexico.
Initiated
Voluntary: Firm initiated
Recall number
D-0773-2026

Terminated Class II Medical devices

Recall started
11 March 2021
Terminated
18 September 2024
Reason
Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than ¿ 35 percent. The dosage displayed on the generator console and table display (monitor) is higher than the actual delivered dose.
Distribution
US Nationwide Distribution and in the country of Canada
Initiated
FDA Mandated
Recall number
Z-1992-2021

Terminated Class II Medical devices

Recall started
27 February 2017
Terminated
16 August 2023
Reason
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
Distribution
US Distribution to: AL, AZ, CA, CT, CO, FL GA, HI, IA, IL, ID, KS, KY, MD, MI, Michigan, MN, NC, NH, NV, NY, OK, OH, PA, SC, TN, TX, VA, WA, WI, and WV,
Initiated
FDA Mandated
Recall numbers
Z-2834-2018, Z-2833-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.