Leica Microsystems, Inc. recalls

15 recalls recorded since 2018, most recently on 21 November 2022. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls15
Ongoing (FDA)3
Terminated or completed12
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
21 November 2022
Reason
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0477-2023, Z-0476-2023, Z-0475-2023

Ongoing Class II Medical devices

Recall started
26 August 2022
Reason
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1850-2022, Z-1848-2022, Z-1849-2022, Z-1851-2022

Terminated Class II Medical devices

Recall started
3 November 2021
Terminated
18 June 2024
Reason
In May 2021, Leica Microsystems received one (1) complaint pertaining to the M220 F12 Surgical Microscope system. There was no patient injury reported for this complaint. However, the identified defect carries some potential injury risk. In the event that this defect may occur, the M220 optics carrier may unintentionally drop into the surgical field, risking contact with the patient.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, DC, FL, GA, IL, IN, MA, MI, MD, NY, OH, OR, PA, TN, TX, VA, WI, and Puerto Rico. The countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Burundi, Canada, Chile, China, C
Initiated
Voluntary: Firm initiated
Recall number
Z-0412-2022

Terminated Class II Medical devices

Recall started
30 July 2021
Terminated
22 June 2023
Reason
Product may not perform as specified in IFU.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2481-2021

Terminated Class II Medical devices

Recall started
20 May 2020
Terminated
13 May 2021
Reason
Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage.
Distribution
Domestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV. International distribution: Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-2399-2020

Ongoing Class II Medical devices

Recall started
18 November 2019
Reason
BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.
Distribution
Nationwide and OUS to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0673-2020

Terminated Class II Medical devices

Recall started
3 May 2019
Terminated
28 September 2023
Reason
The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious injuries.
Distribution
Distributed to Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District Of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska,
Initiated
Voluntary: Firm initiated
Recall number
Z-1729-2019

Terminated Class II Medical devices

Recall started
2 January 2019
Terminated
2 April 2021
Reason
The M220 optics may unintentionally drop into the surgical field, risking contact with the patient.
Distribution
Worldwide distribution: US (nationwide) and countries of: Albania, Algeria, Armenia, Argentina, Bangladesh, Brazil, Burundi, Canada, China, Colombia, Egypt, France, Germany, Guatemala, India, Indonesia, Japan, Nepal, Netherlands, New Zealand, Malaysia, Pakistan, Peru, Philippines, Poland, Qatar, R
Initiated
Voluntary: Firm initiated
Recall number
Z-0772-2019

Terminated Class II Medical devices

Recall started
17 September 2018
Terminated
21 August 2020
Reason
The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.
Distribution
Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0281-2019

Terminated Class II Medical devices

Recall started
4 September 2018
Terminated
10 July 2020
Reason
Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.
Distribution
Distribution in US state of MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-0019-2019

Terminated Class II Medical devices

Recall started
4 September 2018
Terminated
17 August 2021
Reason
The firm received complaints that the instrument tipped over when the swing arm was fully extended.
Distribution
US and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-3175-2018

Terminated Class II Medical devices

Recall started
25 June 2018
Terminated
26 February 2019
Reason
Recall is due to a design weakness of the power insert module.
Distribution
United States distribution: AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, KY, LA, MO, NJ, NY, NV, OH, OR, PA, TX, UT, PR. International distribution to Japan, Brazil, Columbia, Ecuador, Mexico, Taiwan
Initiated
Voluntary: Firm initiated
Recall number
Z-2659-2018

Terminated Class II Medical devices

Recall started
13 June 2018
Terminated
18 March 2019
Reason
These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.
Distribution
Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND
Initiated
Voluntary: Firm initiated
Recall number
Z-0064-2019

Terminated Class II Medical devices

Recall started
4 June 2018
Terminated
25 March 2021
Reason
Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.
Distribution
Worldwide Distribution US Nationwide in the states of : CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX. International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMA
Initiated
Voluntary: Firm initiated
Recall number
Z-2555-2018

Terminated Class II Medical devices

Recall started
25 September 2017
Terminated
21 January 2020
Reason
The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0838-2019