LEICA BIOSYSTEMS NUSSLOCH GMBH recalls

5 recalls recorded since 2018, most recently on 30 January 2025. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)5
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
30 January 2025
Reason
Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
Distribution
Worldwide - US Nationwide distribution in the states of Idaho and Ohio. The countries of France, Australia, Belgium, China, Germany, Spain, UK, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Poland, Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-1247-2025

Ongoing Class II Medical devices

Recall started
4 September 2024
Reason
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0088-2025, Z-0085-2025, Z-0087-2025, Z-0090-2025, Z-0086-2025, Z-0089-2025

Ongoing Class II Medical devices

Recall started
10 May 2024
Reason
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
Distribution
US Nationwide, including government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2529-2024

Ongoing Class II Medical devices

Recall started
10 April 2024
Reason
Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NJ, NV, NY, OH, TN, TX, UT, VA, WA, and WI. There was also government distribution but no military distribution. The country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1947-2024

Ongoing Class II Medical devices

Recall started
20 November 2023
Reason
There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
Distribution
US, Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0590-2024