Laerdal Medical (Suzhou) Co., Ltd. recalls

1 recall recorded since 2018, most recently on 18 December 2020. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
18 December 2020
Terminated
11 December 2023
Reason
Component failures may cause the unit to lose all suction. Users will need to obtain alternate suction devices or methods, leading to a delay in treatment and, in worst case scenario, potential suffocation if alternative suction or ventilation method is not administered by the user.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0868-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.