LABORIE MEDICAL TECHNOLOGIES, CORP recalls

2 recalls recorded since 2018, most recently on 21 May 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
21 May 2026
Reason
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Leban
Initiated
Voluntary: Firm initiated
Recall number
Z-2669-2026

Terminated Class II Medical devices

Recall started
17 July 2019
Terminated
10 July 2023
Reason
Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard
Distribution
US distribution in AZ, and PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2076-2019