LABORIE MEDICAL TECHNOLOGIES, CANADA ULC recalls
1 recall recorded since 2018, most recently on 22 April 2020. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 22 April 2020
- Terminated
- 25 April 2021
- Reason
- This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.
- Distribution
- US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2535-2020
- Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pres