Recall Record

Amsterdam, Netherlands

Kreatech Biotechnology Bv recalls

1 recall recorded since 2018, most recently on 1 November 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
1 November 2024
Reason
The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
Distribution
US States: FL, NY
Initiated
Voluntary: Firm initiated
Recall number
Z-0522-2025

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.