Recall Record

Burlington, MA, United States

Keystone Dental Inc recalls

8 recalls recorded since 2018, most recently on 22 February 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)0
Terminated or completed8
Class I (FDA)0

Terminated Class II Medical devices

Recall started
22 February 2019
Terminated
19 June 2020
Reason
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Distribution
FL, MA, SC, VA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1061-2019, Z-1062-2019, Z-1064-2019, Z-1063-2019

Terminated Class II Medical devices

Recall started
1 May 2018
Terminated
14 September 2018
Reason
The implants are mislabeled.
Distribution
US Distribution to the states of : CA, CT, FL, IN, MA, MD, MI, MN, MT, NJ, NV, NY, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2162-2018

Terminated Class II Medical devices

Recall started
13 March 2018
Terminated
23 July 2018
Reason
The surgical cassette contained extra-large white grommets making the 3.8 mm and 6.5mm dental burrs wobble within the grommets in the surgical cassette.
Distribution
USA (nationwide) Distribution and Internationally to Turkey.
Initiated
Voluntary: Firm initiated
Recall number
Z-1485-2018

Terminated Class II Medical devices

Recall started
24 January 2018
Terminated
30 July 2018
Reason
The cutting edge of tissue punches may be potentially damaged with small dents and nicks.
Distribution
Worldwide Distribution -- US, to the state of California; and, the countries of Canada and Germany
Initiated
Voluntary: Firm initiated
Recall number
Z-1443-2018

Terminated Class III Medical devices

Recall started
16 October 2017
Terminated
17 April 2019
Reason
A locking plastic abutment sleeve, used by dental technicians in dental laboratories to aid in the creation of the final restorative crown, was packed as a non-locking plastic abutment sleeve.
Distribution
International Distribution to: Germany only.
Initiated
Voluntary: Firm initiated
Recall number
Z-1179-2018

Terminated Class II Medical devices

Recall started
25 August 2017
Terminated
16 April 2019
Reason
The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct expiration date should be 2022-06.
Distribution
Nationally.
Initiated
Voluntary: Firm initiated
Recall number
Z-0727-2018

Terminated Class II Medical devices

Recall started
20 June 2017
Terminated
9 April 2018
Reason
Abutments are unable to accept the final abutment screw
Distribution
US Nationwide Distribution in the states of ND, NJ & MA
Initiated
Voluntary: Firm initiated
Recall number
Z-0608-2018

Terminated Class II Medical devices

Recall started
16 June 2017
Terminated
16 April 2019
Reason
An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
Distribution
USA (nationwide) Distribution to the states of : CA, CO, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NY, OH, PA, TN, TX, UT, VA, VT, WA, WI and WV; and Internationally to Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-0718-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.