Johnson & Johnson Surgical Vision, Inc. recalls

15 recalls recorded since 2018, most recently on 31 July 2026. 9 are listed as ongoing. Status and classification are as published by the FDA.

Recalls15
Ongoing (FDA)9
Terminated or completed6
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
31 July 2026
Reason
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
Distribution
Worldwide - US Nationwide distribution including in the states of PA, AL, AZ, AR, CA, FL, IL, IN, IA, KY, LA, ME, MI, MN, MS, NJ, NY, NC, OH, OR, TN, TX, WA, WV, HI, CO, MD, VA, WI, ID, NH, SD, OK, MA, NM, DE, GA, SC, CT, NV, MO, RI, KS, ND and the countries of CA, BE, CH, DE, DK, ES, FR GB, HU, IS,
Initiated
Voluntary: Firm initiated
Recall number
Z-3064-2026

Ongoing Class II Medical devices

Recall started
14 March 2024
Reason
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
Distribution
U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1755-2024

Ongoing Class II Medical devices

Recall started
12 December 2023
Reason
Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
Distribution
International distribution to the country of Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0735-2024

Ongoing Class II Medical devices

Recall started
6 July 2023
Reason
Their is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.
Distribution
International distribution to the countries of China, India, Korea.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2288-2023, Z-2289-2023

Ongoing Class II Medical devices

Recall started
30 May 2023
Reason
Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Distribution
Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chil
Initiated
Voluntary: Firm initiated
Recall number
Z-2247-2023

Ongoing Class II Medical devices

Recall started
7 December 2022
Reason
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Distribution
US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1006-2023, Z-1005-2023, Z-1007-2023

Ongoing Class II Medical devices

Recall started
7 September 2022
Reason
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
Distribution
Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0169-2023

Ongoing Class II Medical devices

Recall started
1 September 2022
Reason
As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.
Distribution
Worldwide Distribution. US states of AL, AR, AZ, CA, CO, DE, FL, GA, HI, IL IN, KY LA, MN, MO, MS, NC, NJ, NY, OH, TN, TX, UT, VA, and WA; Australia, Azerbaijan, Belgium, Brazil, Canada, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Iraq, Ireland, Italy, Japan, Jordan, Kuwa
Initiated
Voluntary: Firm initiated
Recall number
Z-0086-2023

Ongoing Class II Medical devices

Recall started
6 April 2022
Reason
Potential breach in the sterility barrier for tray ring covers.
Distribution
Worlwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, IN, KS, MA, MD, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OR, PA, TN, TX, WA, WI and the countries of Chile, Indonesia, India, Japan, Malaysia, Philippines, Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1033-2022

Terminated Class II Medical devices

Recall started
6 August 2021
Terminated
6 March 2023
Reason
Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.
Distribution
U.S. Nationwide distribution in the state of IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-0028-2022

Terminated Class II Medical devices

Recall started
27 January 2021
Terminated
9 November 2022
Reason
Due to the release of nonconforming Intraocular Lenses (IOLs).
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, FL, GA, IL, LA, MI, MN, NC, NY, OH, PA, SC, TN, TX, UT and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1163-2021

Terminated Class II Medical devices

Recall started
3 December 2019
Terminated
28 April 2021
Reason
Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring additional intervention. Also, reports of clogging of phacoemulsion equipment tubing, which may lead to delay in the procedure or ocular injury.
Distribution
U.S.: CA, PA, NJ, GA, IL, MO, CO, TX, FL, AK, NY, HI, UT, LA, OH, IN, MA, NC, AZ, MI, NH, AL, TN, WA, MD, OR, VA, MS, MN, KY, AR, NV, WI, WV, SC, DE, ME, CT, ID, NM, VT, IA, OK, MT, NE, DC. OUS: DE, GB, DK, IT, NL, TR, AT, CH, FR, GR, ES, PT, FO, BE, GP, ZA, IE, HR, HU, IS, NO, SE, FI
Initiated
Voluntary: Firm initiated
Recall number
Z-0867-2020

Terminated Class II Medical devices

Recall started
24 September 2018
Reason
Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.
Distribution
US Distribution to states of: NC, NV, TX, NY, UT, C, MN, NY, WA, NJ, OH, NV, CA and Internationally to: Australia and Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1893-2019

Terminated Class II Medical devices

Recall started
21 June 2018
Terminated
28 April 2021
Reason
Expired intraocular lenses were distributed.
Distribution
US Nationwide distribution in the states of IN, MN, NC, NV, TX, and WA, WV. There was no foreign/military/government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0464-2019

Terminated Class II Medical devices

Recall started
9 May 2018
Terminated
13 January 2021
Reason
Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
Distribution
US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-0066-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.