Louvain La Neuve, N/A, Belgium

Ion Beam Applications S.A. recalls

26 recalls recorded since 2018, most recently on 3 February 2026. 10 are listed as ongoing. Status and classification are as published by the FDA.

Recalls26
Ongoing (FDA)10
Terminated or completed16
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
3 February 2026
Reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1497-2026

Ongoing Class II Medical devices

Recall started
10 July 2024
Reason
Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
Distribution
IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy Japan Poland Russia Sweden
Initiated
Voluntary: Firm initiated
Recall number
Z-2693-2024

Ongoing Class II Medical devices

Recall started
8 July 2024
Reason
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
Distribution
Worldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL, KA, IL, LA, MI, TN , TX, VA and the countries of BE, CZ, DE, ES, FR, IN, IT, JP, NL, PL, RU, SE, SG, TW.
Initiated
Voluntary: Firm initiated
Recall number
Z-3089-2024

Ongoing Class II Medical devices

Recall started
10 May 2024
Reason
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Distribution
Worldwide - US Nationwide distribution in the state of Virginia and the country of India.
Initiated
Voluntary: Firm initiated
Recall number
Z-1945-2024

Ongoing Class II Medical devices

Recall started
4 March 2024
Reason
There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
Distribution
US Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1603-2024

Ongoing Class II Medical devices

Recall started
18 September 2023
Reason
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
Distribution
Worldwide - US Nationwide distribution in the states of Arkansas, Florida, Kansas, and Virginia and the countries of Belgium, France, India, Italy, Netherlands, Spain, Singapore, and Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0295-2024

Ongoing Class II Medical devices

Recall started
20 April 2022
Reason
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed
Distribution
Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-1205-2022

Ongoing Class II Medical devices

Recall started
15 February 2022
Reason
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
Distribution
IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,
Initiated
Voluntary: Firm initiated
Recall number
Z-0747-2022

Terminated Class II Medical devices

Recall started
29 November 2021
Terminated
20 February 2024
Reason
Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using test processes without notifying the clinical user. If the versions of the processes are different between both environments, this could have an impact on patient treatments.
Distribution
IN USA: VA, OK, NJ, TN OUS: Czech Republic.
Initiated
Voluntary: Firm initiated
Recall number
Z-0466-2022

Ongoing Class II Medical devices

Recall started
14 June 2021
Reason
Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, OK, VA, IL, NJ, TN, WA, LA, TX, MI and the countries of Germany, Italy, Russia, Sweden, India, Japan, France, England, Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2434-2021

Ongoing Class II Medical devices

Recall started
26 May 2021
Reason
if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position
Distribution
FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands
Initiated
Voluntary: Firm initiated
Recall number
Z-2155-2021

Terminated Class II Medical devices

Recall started
17 July 2020
Terminated
30 August 2021
Reason
IBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.
Distribution
US: VA OUS: Belgium
Initiated
Voluntary: Firm initiated
Recall number
Z-2715-2020

Terminated Class II Medical devices

Recall started
21 April 2020
Terminated
1 September 2021
Reason
Signature from the user is necessary to proceed with specific actions in the Proton Therapy System (PTS). IBA became aware that the PTS does not accept user names with more than ten characters. It is an issue when the user has no other choice than resuming an aborted treatment field based on the overall delivered dose displayed on the Dose Counter Electronic Unit (DCEU). User s signature is required to perform this action. If the signature contains more than ten characters, the user will not be able to complete the aborted treatment field. An additional issue applies to Electronic Medical Record (EMR) centric sites and may increase the probability of not being able to complete an aborted treatment field. It is not possible, in a new session, to resume from a local partial archive if the Patient Positioning System (PPS) position has changed. If the user captures the PPS position at every session in the Oncology Information System (OIS) just after the setup process, the prescribed PPS position is changed for the next session in the OIS. This includes the partial continuation session. Therefore, when comparing the prescribed PPS position between OIS and local database, the PTS sees a difference and rejects the local partial archive. This problem forces the user to resume the interrupted irradiation based on the overall delivered dose displayed on the DCEU instead of resuming from the full details of the interrupted beam.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, VA and the countries of Belgium, Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2773-2020

Terminated Class II Medical devices

Recall started
12 December 2018
Terminated
15 April 2020
Reason
There is a risk of detachment of the Compact Gantry Rolling Floor.
Distribution
Worldwide Distribution - US Nationwide in the states of LA and MI. Foreign distribution to France, Japan, and UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-0698-2019

Terminated Class II Medical devices

Recall started
27 June 2018
Terminated
20 May 2020
Reason
Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.
Distribution
US Nationwide Distribution in the states of FL, VA, PA, OK, NJ, WA, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2470-2018

Terminated Class II Medical devices

Recall started
16 May 2018
Terminated
16 October 2018
Reason
Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
Distribution
Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
Initiated
Voluntary: Firm initiated
Recall number
Z-2424-2018

Terminated Class II Medical devices

Recall started
26 January 2018
Terminated
27 April 2020
Reason
When switching from Pencil Beam Scanning treatment mode to Uniform Scanning treatment mode, the tuning setpoints of the scanning magnets are not always taken into account when the proton beam is requested. The mismatch between the setpoint and feedback leads to an error message. Depending on user action, two scenarios are possible: 1. The user resumes the treatment field without analyzing the cause for the error message. This may lead to under-irradiation. 2. The user reboots the Scanning Magnets Power Supply Electronic Unit. This may lead to delay in patient treatment and the patient may need to be re-aligned, which leads to an additional X-ray dose.
Distribution
Consignees in four states: VA, IL, NJ, and WA. One consignee located in Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-1477-2018

Terminated Class II Medical devices

Recall started
5 December 2017
Terminated
29 March 2019
Reason
An accessory (range shifter, ridge filter&) can be improperly inserted in snout 300x400 with one rail only and the proton therapy system can still operate.
Distribution
US Distribution to the states of : MA, FL, VA, PA, OK, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-0934-2018

Terminated Class II Medical devices

Recall started
17 November 2017
Terminated
14 June 2021
Reason
There is a risk of misalignment of the patient when using treatment plans with multiple isocenters.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall number
Z-0938-2018

Terminated Class II Medical devices

Recall started
1 November 2017
Terminated
29 March 2019
Reason
The internal configuration of the electrometers and the Real-Time control boards of the Dekimo Scanning Controller is not checked before each patient treatment.
Distribution
Worldwide Distribution: US (nationwide) to: MI only and country of: France.
Initiated
Voluntary: Firm initiated
Recall number
Z-0943-2018

Terminated Class II Medical devices

Recall started
19 October 2017
Terminated
2 April 2019
Reason
Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.
Distribution
US Distribution to the state of :FL
Initiated
Voluntary: Firm initiated
Recall number
Z-0931-2018

Terminated Class II Medical devices

Recall started
17 October 2017
Terminated
29 March 2019
Reason
IBA identified incorrect screw holes depth in graphite block 8 with respect to the degrader
Distribution
WA
Initiated
Voluntary: Firm initiated
Recall number
Z-0930-2018

Terminated Class II Medical devices

Recall started
28 October 2016
Terminated
9 August 2019
Reason
IBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem.
Distribution
Worldwide Distribution - US Nationwide in the states of LA, TX, Italy, Sweden and France. Each country's National Competent Authorities were notified of the firm's Field Safety Notice.
Initiated
Voluntary: Firm initiated
Recall number
Z-2109-2018

Terminated Class II Medical devices

Recall started
16 September 2016
Terminated
8 February 2019
Reason
IBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk.
Distribution
Units were delivered in the USA, Korea, Germany, France, Czech Republic, Italy, Poland and Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2139-2018, Z-2138-2018

Terminated Class II Medical devices

Recall started
7 August 2016
Terminated
25 March 2020
Reason
IBA is recalling to provide information through a Safety Notice to reduce risk for maintenance technicians during ion source replacement.
Distribution
Units were distributed in the USA, Korea, Germany, France, Czech Republic, Italy, Poland, and Sweden. Each country's national Competent Authorities have been informed.
Initiated
Voluntary: Firm initiated
Recall number
Z-2173-2018

Terminated Class II Medical devices

Recall started
11 April 2016
Terminated
25 March 2020
Reason
IBA is conducting a voluntary recall to address a PTS (Proton Therapy System) software issue and to reduce the risk related to this issue.
Distribution
Units were distributed to Korea and Jacksonville, FL.
Recall number
Z-2284-2018