Orlando, FL, United States

Invivo Corporation recalls

7 recalls recorded since 2018, most recently on 14 September 2020. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)1
Terminated or completed6
Class I (FDA)0

Terminated Class III Medical devices

Recall started
14 September 2020
Terminated
14 August 2023
Reason
The labels have been mistakenly printed with an IPX1 rating symbol. IPX ratings relate to protections that products have from water ingress.
Distribution
US Nationwide Distribution: AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OK, OH, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, and WV; and the countries of: Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Banglade
Initiated
Voluntary: Firm initiated
Recall number
Z-0126-2021

Terminated Class II Medical devices

Recall started
6 February 2020
Terminated
4 May 2022
Reason
Insufficient shielding of Philips SensaVue DVI Adapter for computer monitor.
Distribution
U.S Distribution: AK, AL, CA, CO, FL, IL, IN, KY, MA, MI, NY, OH, OR, PA, TN, TX and WV. *** Foreign: Australia, Canada, France, United Kingdom, Italy, Japan, Netherlands, Saudi Arabia, India and Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1440-2020

Completed Class II Medical devices

Recall started
24 September 2018
Reason
The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard
Distribution
Distributed to the Netherlands
Initiated
Voluntary: Firm initiated
Recall number
Z-0462-2019

Terminated Class II Medical devices

Recall started
3 August 2018
Terminated
15 July 2021
Reason
Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.
Distribution
Distribution in the Netherlands
Initiated
Voluntary: Firm initiated
Recall number
Z-3192-2018

Terminated Class II Medical devices

Recall started
14 March 2018
Terminated
30 April 2021
Reason
The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
Distribution
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico., and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Kazakhstan, K
Initiated
Voluntary: Firm initiated
Recall number
Z-1867-2018

Terminated Class II Medical devices

Recall started
24 January 2018
Terminated
4 May 2020
Reason
Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers and could result in visually underestimating calculated Ktrans, Kep, and iAUGC values. The defect also impacts DynaCAD s on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps. If the ADC values are computed on-the-fly by DynaCAD, the ADC values and colors will also display incorrectly if viewed on a remote DynaCAD client.
Distribution
Device is software. Customer notification letters recommended that users discontinue use of the Ktrans, Kep, and iAUGC colormaps when assessing studies from a remotely connected DynaCAD client computer. It is further advised that users refrain from using the colormaps and values derived from ADC ma
Initiated
Voluntary: Firm initiated
Recall number
Z-1289-2018

Ongoing Class II Medical devices

Recall started
20 November 2017
Reason
While in Full Disclosure playback, a user may inadvertently close the Full Disclosure Control Window using the ESC key function, instead of pressing the X icon in the upper right corner of this window.
Distribution
Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, NH, NJ, NM, NY, NV, OH, PA, RI, SC, TN, TX, VA, WA, and WI ; and countries of: Canada, China, Great Britain and Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-2115-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.