Recall Record

Elyria, OH, United States

Invacare Corporation recalls

7 recalls recorded since 2018, most recently on 22 January 2024. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)5
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
22 January 2024
Reason
Invacare Homecare & G-Series Bed Components identified with a potential weld defect leads to premature weld failures to occur on these components
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1251-2024

Ongoing Class II Medical devices

Recall started
15 June 2023
Reason
Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2445-2023

Ongoing Class III Medical devices

Recall started
10 June 2022
Reason
Between March 2, 2021 and December 31, 2021, the product label used in production did not contain the "Rx Only" statement as required.
Distribution
Domestic distribution to the following states: ALASKA ARIZONA CALIFORNIA COLORADO GEORGIA IDAHO IOWA KANSAS KENTUCKY MICHIGAN MINNESOTA MISSOURI NEW YORK NORTH CAROLINA OHIO OKLAHOMA PENNSYLVANIA TENNESSEE
Initiated
Voluntary: Firm initiated
Recall number
Z-1369-2022

Ongoing Class II Medical devices

Recall started
30 March 2022
Reason
The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1130-2022

Ongoing Class II Medical devices

Recall started
12 November 2021
Reason
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0446-2022, Z-0445-2022, Z-0447-2022

Terminated Class II Medical devices

Recall started
30 October 2020
Terminated
31 December 2024
Reason
Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.
Distribution
Worldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands
Initiated
Voluntary: Firm initiated
Recall number
Z-0514-2021

Terminated Class II Medical devices

Recall started
16 November 2018
Terminated
15 April 2020
Reason
Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained. Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur. For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures.
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Canada, Colombia, Germany, Guam, India, Northern Mariana Islands, Philippines and South Africa.
Initiated
Voluntary: Firm initiated
Recall number
Z-0549-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.