Intuitive Surgical, Inc. recalls

53 recalls recorded since 2018, most recently on 24 April 2026. 35 are listed as ongoing. Status and classification are as published by the FDA.

Recalls53
Ongoing (FDA)35
Terminated or completed18
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
24 April 2026
Reason
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Belgium, France, Germany, Italy, Japan, Netherlands, New Zealand, Poland,
Initiated
Voluntary: Firm initiated
Recall number
Z-2310-2026

Ongoing Class II Medical devices

Recall started
22 April 2026
Reason
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
Distribution
Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Swit
Initiated
Voluntary: Firm initiated
Recall number
Z-2521-2026

Ongoing Class II Medical devices

Recall started
2 April 2026
Reason
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Au
Initiated
Voluntary: Firm initiated
Recall number
Z-2046-2026

Ongoing Class I Medical devices

Recall started
11 March 2026
Reason
Reloads for curved-tipped stapler may potentially produce an incomplete staple line.
Distribution
US: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware District of Columbia Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New Mex
Initiated
Voluntary: Firm initiated
Recall number
Z-1855-2026

Ongoing Class II Medical devices

Recall started
30 December 2025
Reason
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
Distribution
US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1355-2026

Ongoing Class II Medical devices

Recall started
15 December 2025
Reason
Access Port System tray may develop cracks potentially resulting in a sterility breach.
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nevada, New Hampshire, New Jersey
Initiated
Voluntary: Firm initiated
Recall number
Z-1096-2026

Ongoing Class II Medical devices

Recall started
9 December 2025
Reason
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
Distribution
U.S. Nationwide distribution in the states of AR, CA, CT, FL, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, NC, OH, PA, TX, VA, and WA. The countries of Argentina, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Gua
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1701-2026, Z-1700-2026

Ongoing Class II Medical devices

Recall started
9 December 2025
Reason
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Distribution
Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1698-2026, Z-1699-2026, Z-1696-2026, Z-1697-2026

Ongoing Class II Medical devices

Recall started
4 December 2025
Reason
Due to probe bags that may have a compromised or incomplete sterile pouch seal.
Distribution
U.S.: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MA, MD, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. O.U.S.: Australia, China, France, Germany, Italy, Japan, South Korea, Spain, Switzerl
Initiated
Voluntary: Firm initiated
Recall number
Z-1134-2026

Ongoing Class II Medical devices

Recall started
17 November 2025
Reason
An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.
Distribution
US distribution: LA, MA, NY, PA, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1428-2026

Ongoing Class II Medical devices

Recall started
7 November 2025
Reason
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
Distribution
U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Korea, Spain, Switzerland, Taiwan and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-1232-2026

Ongoing Class II Medical devices

Recall started
12 May 2025
Reason
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
Distribution
Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, J
Initiated
Voluntary: Firm initiated
Recall number
Z-2722-2026

Ongoing Class II Medical devices

Recall started
26 February 2025
Reason
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
Distribution
Worldwide distribution - US Nationwide and the countries of Alaska, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebra
Initiated
Voluntary: Firm initiated
Recall number
Z-1325-2025

Ongoing Class II Medical devices

Recall started
19 December 2024
Reason
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Distribution
U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebra
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0870-2025, Z-0880-2025, Z-0849-2025, Z-0874-2025, Z-0847-2025, Z-0862-2025, Z-0859-2025, Z-0872-2025, Z-0851-2025, Z-0857-2025, Z-0858-2025, Z-0873-2025 and 25 more

Ongoing Class II Medical devices

Recall started
19 December 2024
Reason
Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments
Distribution
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY. The countries of Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0924-2025, Z-0923-2025

Ongoing Class II Medical devices

Recall started
21 August 2024
Reason
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
Distribution
US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
Initiated
Voluntary: Firm initiated
Recall number
Z-0964-2026

Ongoing Class II Medical devices

Recall started
12 July 2024
Reason
There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
Distribution
US Nationwide distribution in the states of AZ and CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2720-2024

Ongoing Class II Medical devices

Recall started
13 March 2024
Reason
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
Distribution
US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-1518-2024

Ongoing Class II Medical devices

Recall started
6 October 2023
Reason
There is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.
Distribution
US: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-0406-2024

Ongoing Class II Medical devices

Recall started
12 June 2023
Reason
Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.
Distribution
Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ, OH, IL, NC, GA, KS, CA, HI, MD, OR, SC, AK, NY, NE, IA, WA, OK, ND, UT, MS, AR, WI, MI, TX, TN, IN, NJ, VA, SD, CO, KY, AL, MO, LA, DE, MA, MT, WV, ID, CT, NV, NH, ME, WY, DC, RI, NM, VT and the countries of Australia, Austr
Initiated
Voluntary: Firm initiated
Recall number
Z-2059-2023

Ongoing Class II Medical devices

Recall started
10 February 2023
Reason
Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail component not being manufactured within specifications.
Distribution
U.S.: Cayman Islands, Dominican Republic, Puerto Rico, United States O.U.S.: Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Luxembourg, Martinique, Monaco, Netherlands, Norway, Reunion, Sweden, Switzerland, United Kingdom, India, Japan, South Korea, Taiwan, Brazil, Argentina, Austral
Initiated
Voluntary: Firm initiated
Recall number
Z-1293-2023

Ongoing Class II Medical devices

Recall started
24 January 2023
Reason
Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.
Distribution
Worldwide - US Nationwide distribution in the states of FL, GA, WA and the countries of United Kingdom, Germany, France
Initiated
Voluntary: Firm initiated
Recall number
Z-1262-2023

Ongoing Class II Medical devices

Recall started
7 October 2022
Reason
Potential staple deployment failure and device fragment generation
Distribution
U.S.: Arizona, California, Colorado, Florida, Illinois, Indiana, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Nevada, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, and Wisconsin. O.U.S.: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0340-2023

Ongoing Class II Medical devices

Recall started
20 September 2022
Reason
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Distribution
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Initiated
Voluntary: Firm initiated
Recall number
Z-0133-2023

Ongoing Class II Medical devices

Recall started
1 July 2022
Reason
Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The coun
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1459-2022, Z-1460-2022

Completed Class II Medical devices

Recall started
22 June 2022
Reason
Due to preventative maintenance (PM) not being performed on E-100 generators which is a bipolar electrosugical unit designed to be used with the da Vinci Xi and X systems.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY. The countries of Belgium, Denmark, Germany,
Initiated
Voluntary: Firm initiated
Recall number
Z-1484-2022

Ongoing Class II Medical devices

Recall started
17 March 2022
Reason
As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with bleeding of critical severity exceeds the predicted likelihood of harm as documented within Intuitive s existing risk documentation.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The coun
Initiated
Voluntary: Firm initiated
Recall number
Z-0999-2022

Completed Class II Medical devices

Recall started
3 March 2022
Reason
Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result in an insufficient seal.
Distribution
Worldwide distribution - US Nationwide and the countries/regions of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Leb
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1209-2022, Z-1210-2022

Ongoing Class II Medical devices

Recall started
23 December 2021
Reason
Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.
Distribution
Worldwide distribution - US Nationwide distribution in the states of California, Wisconsin, Virginia, Ohio, Texas, New York, Nevada, Connecticut, New Jersey, Massachusetts, Georgia, Florida, Indiana, North Carolina, Missouri, Pennsylvania, Michigan, Oregon, Kentucky, Hawaii, Tennessee, Alabama, Nebr
Initiated
Voluntary: Firm initiated
Recall number
Z-0702-2022

Ongoing Class II Medical devices

Recall started
22 September 2021
Reason
Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.
Distribution
US Nationwide distribution in the states of Arizona, California, Connecticut, Florida, Illinois, Kansas, Kentucky, Massachusetts, Minnesota, Missouri, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington, and Wisconsin.
Initiated
Voluntary: Firm initiated
Recall number
Z-0419-2022

Terminated Class II Medical devices

Recall started
21 September 2021
Terminated
25 January 2024
Reason
Incorrect expiration date on its package labeling (shipper box and inner carton label).
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, CA, GA, IL, FL, CO, TX, MO, OH, MA, WI, NC, LA, UT and the country of South Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-0275-2022

Terminated Class II Medical devices

Recall started
6 August 2021
Terminated
17 May 2024
Reason
Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.
Distribution
US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY, NC, OH, OK, PA, TN, TX, UT, VA, WA OUS: China
Initiated
Voluntary: Firm initiated
Recall number
Z-0106-2022

Terminated Class II Medical devices

Recall started
23 October 2020
Terminated
5 December 2022
Reason
The distal tip ring of the fully articulating catheter may become dislodged during the procedure.
Distribution
US Nationwide distribution including in the states of CA, CO, DC, FL, IA, IL, MA, MD, MI, MN, NC, NJ, NY, OH, PA, TX, UT, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0489-2021

Ongoing Class II Medical devices

Recall started
10 June 2020
Reason
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Distribution
U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2949-2020, Z-2950-2020

Terminated Class II Medical devices

Recall started
19 December 2019
Terminated
31 August 2023
Reason
Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Arkansas, Arizona, Alabama, Alaska, California, Connecticut, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Idaho, Kentucky, Kansas, Louisiana, Missouri, Michigan, Maryland, Mississippi, Minnesota, Montana,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2328-2020, Z-2329-2020

Ongoing Class II Medical devices

Recall started
19 August 2019
Reason
Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
Distribution
Distribution US nationwide and South Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-2131-2020

Terminated Class II Medical devices

Recall started
26 July 2019
Terminated
10 August 2022
Reason
The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Distribution
Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Massachusetts, Michigan, Minnesota, Missouri, Montana, New Jersey, New Mexico, New York, North Caroli
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0926-2020, Z-0925-2020, Z-0927-2020

Ongoing Class II Medical devices

Recall started
1 July 2019
Reason
The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.
Distribution
Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denma
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1891-2020, Z-1890-2020

Terminated Class II Medical devices

Recall started
17 June 2019
Terminated
18 July 2022
Reason
One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
Distribution
US Nationwide distribution including in the state of Washington.
Initiated
Voluntary: Firm initiated
Recall number
Z-0283-2021

Terminated Class II Medical devices

Recall started
11 April 2019
Terminated
8 April 2024
Reason
Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1897-2020, Z-1896-2020

Terminated Class II Medical devices

Recall started
1 April 2019
Terminated
16 December 2021
Reason
Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following, extremely rare specific conditions: 1) the user is activating Adjust Mode, AND 2) the instrument Arm is near its vertical position limit, AND 3) the user is applying sustained force against the hand control haptic feedback.
Distribution
U.S.: OH, PN, NY, MN, TN, TX, OR, IL, MD, NJ, AL. O.U.S.: South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-2260-2020

Ongoing Class II Medical devices

Recall started
16 January 2019
Reason
During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.
Distribution
US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United
Initiated
Voluntary: Firm initiated
Recall number
Z-0841-2019

Terminated Class II Medical devices

Recall started
4 December 2018
Terminated
24 August 2022
Reason
This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
Distribution
US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, No
Initiated
Voluntary: Firm initiated
Recall number
Z-1105-2020

Terminated Class II Medical devices

Recall started
21 November 2018
Terminated
31 May 2022
Reason
Due to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error 23087) and loss of arm functionality prior or during surgery.
Distribution
U.S.: CA, MD, NY, VA, IL, MI, DE, TX, FL, NJ, MA, KY and OH. O.U.S.: Australia, France, Germany and Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1960-2020, Z-1961-2020

Terminated Class II Medical devices

Recall started
25 October 2018
Terminated
5 April 2024
Reason
The instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier.
Distribution
Updated information provided on 03/18/2020: FL, OH, TN, NJ, PA, MD, NY, CA, MN, NV, OR, HI, VA, TX, AL, CA, IL, CN, and NC. OUS: APAC (Asia Pacific) - South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-1906-2020

Terminated Class II Medical devices

Recall started
23 October 2018
Terminated
21 January 2022
Reason
The firm became aware of a problem with the surgical system where the cannula mount button may be damaged when the user releases the button abruptly.
Distribution
Product distributed in the US to the following states: Minnesota, New York, Ohio, and Pennsylvania.
Initiated
Voluntary: Firm initiated
Recall number
Z-0845-2020

Ongoing Class II Medical devices

Recall started
15 August 2018
Reason
Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.
Distribution
Worldwide Distribution - US Nationwide in the states of : AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MA, ME, MD, MI, MN, MS, MO, NB, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY and the countries of : Argentina, Aus
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3024-2018, Z-3025-2018

Terminated Class II Medical devices

Recall started
16 July 2018
Terminated
5 October 2023
Reason
Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patient Side Carts were manufactured incorrectly and may fail prior or during use.
Distribution
Worldwide Distribution: US (nationwide) and countries of: Austria, Australia, Belgium, Brazil, Canada, Switzerland, China, Czech Republic, Germany, Denmark, Ecuador, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Ne
Initiated
Voluntary: Firm initiated
Recall number
Z-3254-2018

Ongoing Class II Medical devices

Recall started
4 June 2018
Reason
The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation.
Distribution
Worldwide distribution. US nationwide, Australia, Italy, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0249-2020

Ongoing Class II Medical devices

Recall started
17 May 2018
Reason
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Distribution
U.S.
Initiated
Voluntary: Firm initiated
Recall number
Z-2835-2018

Terminated Class II Medical devices

Recall started
23 April 2018
Terminated
15 July 2020
Reason
Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.
Distribution
U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2381-2018, Z-2380-2018

Terminated Class II Medical devices

Recall started
7 March 2018
Terminated
29 July 2020
Reason
Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
Distribution
US Nationwide AR, CA, FL, GA, IL, IA, MN, MS, NE, OK, TN, and TX. Foreign distribution was made to Australia, Austria, Finland, France, Germany, Italy, Netherlands, Spain, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2341-2018

Terminated Class II Medical devices

Recall started
19 September 2017
Terminated
30 January 2024
Reason
Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0404-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.