Instrumentation Laboratory recalls

18 recalls recorded since 2018, most recently on 3 September 2026. 15 are listed as ongoing. Status and classification are as published by the FDA.

Recalls18
Ongoing (FDA)15
Terminated or completed3
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
3 September 2026
Reason
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Kore
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0002-2027, Z-0001-2027

Ongoing Class II Medical devices

Recall started
21 May 2026
Reason
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2900-2026, Z-2899-2026

Ongoing Class II Medical devices

Recall started
8 May 2026
Reason
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2825-2026, Z-2823-2026, Z-2824-2026, Z-2822-2026, Z-2821-2026, Z-2826-2026

Ongoing Class II Medical devices

Recall started
20 January 2026
Reason
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1524-2026, Z-1533-2026, Z-1539-2026, Z-1538-2026, Z-1534-2026, Z-1531-2026, Z-1527-2026, Z-1529-2026, Z-1542-2026, Z-1530-2026, Z-1525-2026, Z-1541-2026 and 8 more

Ongoing Class II Medical devices

Recall started
11 December 2025
Reason
Potential for microbial contamination.
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, MN, MS, NE, OH, TX, WI and the countries of Belgium, Netherlands, Luxembourg, Switzerland, China, Hong Kong, Israel, Italy, South Korea, Lithuania, Macao, Malaysia, Philippines, Qatar, Romania, Turkey, Kosovo.
Initiated
Voluntary: Firm initiated
Recall number
Z-1095-2026

Ongoing Class III Medical devices

Recall started
18 November 2025
Reason
Recalled lots were manufactured with double the amount of preservative concentration.
Distribution
Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
Initiated
Voluntary: Firm initiated
Recall number
Z-0938-2026

Ongoing Class II Medical devices

Recall started
3 February 2025
Reason
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Distribution
Worldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, J
Initiated
Voluntary: Firm initiated
Recall number
Z-2323-2025

Ongoing Class II Medical devices

Recall started
18 December 2024
Reason
Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.
Distribution
Worldwide - US Nationwide and the countries of Andorra, Algeria, Armenia, Australia, Austria, Belarus, Belgium, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Cyprus, Czechia, Denmark, Finland, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iran (Islamic Repub
Initiated
Voluntary: Firm initiated
Recall number
Z-1020-2025

Ongoing Class II Medical devices

Recall started
29 September 2023
Reason
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Distribution
US States: MA, MN, NY. India.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0358-2024, Z-0357-2024, Z-0359-2024

Ongoing Class II Medical devices

Recall started
13 July 2023
Reason
Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Finland, Ireland, Lithuania, Malta, Qatar, Sweden.
Initiated
Voluntary: Firm initiated
Recall number
Z-2578-2023

Ongoing Class II Medical devices

Recall started
3 February 2023
Reason
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Albania, Algeria, Australia, Austria, Azerbaijan, Belarus, Belgium, Brazil, Bulgaria, Cape Verde, China, Colombia, Czech Republic, Egypt, Falkland Islands, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, Ind
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1306-2023, Z-1305-2023, Z-1304-2023, Z-1307-2023

Ongoing Class II Medical devices

Recall started
8 April 2022
Reason
Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
Distribution
US Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1052-2022

Ongoing Class II Medical devices

Recall started
22 February 2022
Reason
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, H
Initiated
Voluntary: Firm initiated
Recall number
Z-1014-2022

Ongoing Class II Medical devices

Recall started
13 October 2021
Reason
The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion includes states of IN, MN, and NC.
Initiated
Voluntary: Firm initiated
Recall number
Z-0276-2022

Ongoing Class II Medical devices

Recall started
25 August 2021
Reason
Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days
Distribution
US Nationwide Foreign: Canada Armenia Austria Australia Belgium Brunei Brazil Belarus Switzerland China Colombia Czech Republic Germany Denmark Spain Finland France United Kingdom
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0151-2022, Z-0150-2022, Z-0149-2022

Terminated Class II Medical devices

Recall started
23 July 2019
Terminated
27 April 2020
Reason
Prolonged clotting times
Distribution
States: IL WI MI Canada, Spain, Portugal, Poland, Italy, Israel, Switzerland, Denmark, Kazakhstan, United Kingdom, Germany, Netherlands, Belgium, France, Jordan, India, Hong Kong, Austria, Czechia, Australia, Sri Lanka, Argentina, Vietnam, Singapore, Japan, South Korea, Cyprus, Ireland, Hungar
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2486-2019, Z-2487-2019, Z-2485-2019

Terminated Class II Medical devices

Recall started
1 August 2018
Terminated
16 April 2019
Reason
There is a potential for increased imprecision, out of range quality controls and prolonged sample results.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0077-2019

Terminated Class II Medical devices

Recall started
2 August 2017
Terminated
29 March 2019
Reason
Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.
Distribution
Worldwide Distribution - US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0895-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.