Inpeco S.A. recalls
14 recalls recorded since 2018, most recently on 30 October 2025. 7 are listed as ongoing. Status and classification are as published by the FDA.
Recalls14
Ongoing (FDA)7
Terminated or completed7
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 30 October 2025
- Reason
- The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
- Distribution
- US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0596-2026, Z-0595-2026
- FlexLab (FLX); Version: FLX-217-40;
- FlexLab (FLX); Version: FLX-217-10;
Ongoing Class II Medical devices
- Recall started
- 3 September 2025
- Reason
- The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
- Distribution
- US distribution to CA & NY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0107-2026, Z-0106-2026
- FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
- FlexLab (FLX) System. Potassium Test System. in vitro diagnostic
Ongoing Class II Medical devices
- Recall started
- 3 November 2023
- Reason
- Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
- Distribution
- Worldwide - US Nationwide distribution in the state of MA and the countries of AUSTRIA, BELGIUM, DENMARK, ITALY, NORWAY, SPAIN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0560-2024
- Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample pr
Ongoing Class II Medical devices
- Recall started
- 10 March 2023
- Reason
- Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Distribution
- Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EAU, EGYPT, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, MALAYSIA, MEXICO, MYANMAR, NETHERLANDS, NO
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1560-2023, Z-1559-2023, Z-1562-2023, Z-1561-2023, Z-1563-2023
- Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: ACP-207-00, ACP-207-01, ACP-230-00, A
- Input Output Module (IOM) FLX-201-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-201-00, APT-201-00, APT-253-00 A
- Vesmatic Cube 80 Interface Module (VMC)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: 72747000.A
- PVT Interface Module (PVT)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: 72747000.A
- Alinity h Interface Module (HSQ)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-274-20
Ongoing Class II Medical devices
- Recall started
- 10 September 2021
- Reason
- The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
- Distribution
- Domestic distribution nationwide . Foreign distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0087-2022
- FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)
Terminated Class II Medical devices
- Recall started
- 7 August 2020
- Terminated
- 2 December 2021
- Reason
- Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2974-2020, Z-2975-2020, Z-2973-2020
- Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Aut
- Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Aut
- FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory Automation system - Product Usage: The FlexLab 3.6 Automation is
Terminated Class II Medical devices
- Recall started
- 14 July 2020
- Terminated
- 28 December 2023
- Reason
- When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
- Distribution
- Global Distribution including: USA and countries of: Algeria, Australia, Azerbaijan, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Egypt, Finland, France, Germany, Indonesia, Israel, Italy, Libano, Mexico, Netherlands, Philippines, Portugal, Qatar, Saudi Arabia, Singapore, South Africa,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0433-2021, Z-0434-2021, Z-0432-2021
- Accelerator a3600 Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical
- Aptio Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing,
- FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical proce
Ongoing Class II Medical devices
- Recall started
- 7 April 2020
- Reason
- The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
- Distribution
- Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1888-2020, Z-1889-2020, Z-1887-2020
- Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
- Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
- Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
Terminated Class II Medical devices
- Recall started
- 7 April 2020
- Terminated
- 28 July 2022
- Reason
- When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
- Distribution
- Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansa
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2003-2020, Z-2004-2020
- Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-
- Abbott Accelerator a3600, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: ACP-208-00 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-An
Terminated Class II Medical devices
- Recall started
- 7 April 2020
- Terminated
- 6 February 2024
- Reason
- The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
- Distribution
- US Nationwide including in the states of IL, NY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2141-2020, Z-2140-2020, Z-2142-2020
- Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) is a modula
- FlexLab Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00)is a modular system de
- Aptio Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) - Product Usage: is a m
Ongoing Class II Medical devices
- Recall started
- 7 April 2020
- Reason
- In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
- Distribution
- US Nationwide distribution including in the states of IL, NY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2199-2020, Z-2200-2020, Z-2198-2020
- Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.
- Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.
- FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.
Terminated Class II Medical devices
- Recall started
- 7 April 2020
- Terminated
- 13 June 2024
- Reason
- The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
- Distribution
- The products were distributed to the following US States: IL and NY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2114-2020, Z-2115-2020
- FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM4
- Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output M
Terminated Class II Medical devices
- Recall started
- 24 August 2018
- Terminated
- 19 March 2020
- Reason
- The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes.
- Distribution
- Foreign distribution to Netherlands, Italy, Belgium, Austria, France, Spain, and Germany.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0361-2019
- FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre- Analytical and Post-Analytical processing,
Terminated Class II Medical devices
- Recall started
- 13 February 2018
- Terminated
- 3 April 2019
- Reason
- Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
- Distribution
- Distributors in 2 states: NY and IL.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1798-2018
- FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.