In2Bones, SAS recalls

2 recalls recorded since 2018, most recently on 26 January 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
26 January 2024
Reason
This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
Distribution
US Nationwide distribution in the state of Tennessee.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1320-2024, Z-1319-2024

Terminated Class II Medical devices

Recall started
9 February 2021
Terminated
11 July 2024
Reason
This Field Action is being conducted following identification of manufacturing non-conformity of some batches of Shannon burrs diameter 2.0mm length 20mm, possibly causing the breakage of the burr during use.
Distribution
US Nationwide distribution in the state of TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1283-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.