IMPLANTCAST GMBH recalls

2 recalls recorded since 2018, most recently on 10 May 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
10 May 2021
Terminated
18 November 2022
Reason
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.
Distribution
U.S. Nationwide distribution in the states of FL and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2050-2021

Terminated Class II Medical devices

Recall started
28 December 2017
Terminated
1 September 2020
Reason
Possibility that the retaining ring in the 2M inlay 15degree for MUTARS RS Cup and LUMiC TiN is incorrectly mounted.
Distribution
Distribution in Europe and US states of Texas, Rhode Island and Georgia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2517-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.