ImaCor Inc. recalls

2 recalls recorded since 2018, most recently on 29 August 2025. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)2
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
29 August 2025
Reason
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
Distribution
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0038-2026

Ongoing Class II Medical devices

Recall started
16 September 2024
Reason
Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.
Distribution
FL ,GA, NJ, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-0231-2025