Illumina, Inc. recalls
5 recalls recorded since 2018, most recently on 28 October 2025. 2 are listed as ongoing. Status and classification are as published by the FDA.
Recalls5
Ongoing (FDA)2
Terminated or completed3
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 28 October 2025
- Reason
- An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.
- Distribution
- Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT, IL, MD, NY, VA, WV, SD, TX, AL, SC, TN, KY, MI and the countries of DK, SA, FR, DE, IT, CN, PL, GB, TH, GR, VN, BE, ES, IL, KR, AU, CH, BG, NL, IE, TW, TR, CZ, HU, NO, FI, JP, SG, LU, AT, AE, ZA, PT, LT, SE.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0884-2026
- Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles, REF: 20031098; NextSeq 55
Ongoing Class II Medical devices
- Recall started
- 17 September 2024
- Reason
- Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, KS, NC, CA, UT, KA, MA, NJ, MD and the countries of Spain, South Korea, Japan, Australia, Italy, United Arab Emirates, Sweden, United Kingdom, Czech Republic, Singapore, Austria, Poland, France, Germany, Finland, Switzerland, Slovakia.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0068-2027
- NovaSeq 6000Dx including DRAGEN Server, Catalog Number: 20068232; Illumina DRAGEN Server for NextSeq 550Dx, Catalog Number: 20086130
Completed Class II Medical devices
- Recall started
- 5 April 2023
- Reason
- Cybersecurity vulnerability concerning the software used for sequencing instruments.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, AU Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Egypt, FR France, GB Georgia, Germany, Guatemala, Hong Kong, Hungary, IT Italy, Japan, Latvia, Lithuania, Luxembourg, Macedonia, Malt
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1977-2023, Z-1976-2023
- MiSeq Dx Instrument, Catalog # M70101, containing Universal Copy Service (UCS) software.
- NextSeq 550Dx Instrument, Catalog # 20005715, contaning Universal Copy Service software
Completed Class II Medical devices
- Recall started
- 3 May 2022
- Reason
- cybersecurity vulnerability
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, and WV. The countries of Algeria, Australia, Austri
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1260-2022, Z-1259-2022
- illumina Model NextSeq 550 Dx REF 20005715
- illumina REF DX-410-1001 Model: MiSeq Dx illumina REF 15036706 Model: MiSeq Dx illumina REF 20014053 Model: MiSeq Dx
Completed Class II Medical devices
- Recall started
- 22 February 2022
- Reason
- There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.
- Distribution
- US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federatio
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0854-2022
- Illumina NextSeq 550Dx, REF: 20005715, CE, IVD
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.