Ongoing Class II Medical devices

Recall started
28 November 2023
Reason
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Distribution
US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0700-2024, Z-0701-2024

Terminated Class II Medical devices

Recall started
3 November 2022
Terminated
10 September 2026
Reason
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Distribution
US Nationwide distribution in the state of GA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0728-2023