Howmedica Osteonics Corp. recalls

48 recalls recorded since 2018, most recently on 16 September 2026. 26 are listed as ongoing. Status and classification are as published by the FDA.

Recalls48
Ongoing (FDA)26
Terminated or completed22
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
16 September 2026
Reason
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0164-2027, Z-0166-2027, Z-0165-2027, Z-0167-2027

Ongoing Class II Medical devices

Recall started
30 July 2026
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3206-2026, Z-3199-2026, Z-3201-2026, Z-3196-2026, Z-3204-2026, Z-3197-2026, Z-3200-2026, Z-3205-2026, Z-3202-2026, Z-3203-2026, Z-3198-2026, Z-3195-2026

Ongoing Class II Medical devices

Recall started
19 May 2026
Reason
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
Distribution
US Nationwide distribution in the states of AZ, CA, GA, MI, NC, NJ, NY, WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2461-2026, Z-2462-2026

Ongoing Class II Medical devices

Recall started
2 February 2026
Reason
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Initiated
Voluntary: Firm initiated
Recall number
Z-1572-2026

Ongoing Class II Medical devices

Recall started
12 November 2025
Reason
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-0963-2026

Ongoing Class II Medical devices

Recall started
28 August 2025
Reason
a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
Distribution
International distribution in the country of United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2640-2025

Ongoing Class II Medical devices

Recall started
4 September 2024
Reason
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
Distribution
US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of t
Initiated
Voluntary: Firm initiated
Recall number
Z-0023-2025

Ongoing Class II Medical devices

Recall started
7 August 2024
Reason
Stryker has identified an issue with the characterization process that impacted the constant values assigned to the associated MICS Handpieces.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-0044-2025

Ongoing Class II Medical devices

Recall started
5 August 2024
Reason
A patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.
Distribution
International distribution to the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2732-2024

Ongoing Class II Medical devices

Recall started
19 June 2024
Reason
Mislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, and UK.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2370-2024, Z-2372-2024, Z-2371-2024

Ongoing Class II Medical devices

Recall started
23 May 2024
Reason
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2153-2024, Z-2152-2024

Ongoing Class II Medical devices

Recall started
27 February 2024
Reason
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
Distribution
Nationwide Foreign: Netherlands, Canada, Singapore, Spain, Sweden, UK
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1631-2024, Z-1628-2024, Z-1627-2024, Z-1622-2024, Z-1630-2024, Z-1632-2024, Z-1623-2024, Z-1624-2024, Z-1626-2024, Z-1629-2024, Z-1621-2024, Z-1625-2024

Ongoing Class II Medical devices

Recall started
6 February 2024
Reason
Potential packaging breaches of inner blister and outer sterile blister.
Distribution
Domestic: Nationwide Distribution; Foreign: Australia, Canada, Chile, Colombia, Korea, & Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1264-2024, Z-1260-2024, Z-1259-2024, Z-1263-2024, Z-1262-2024, Z-1261-2024, Z-1266-2024, Z-1265-2024

Ongoing Class II Medical devices

Recall started
22 December 2023
Reason
Failed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, China (Hong Kong), Columbia, Italy, Japan, Malaysia, Netherlands (Venlo), Slovenia, S. Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0999-2024, Z-1000-2024

Ongoing Class II Medical devices

Recall started
1 December 2023
Reason
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Distribution
Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0885-2024, Z-0914-2024, Z-0923-2024, Z-0886-2024, Z-0910-2024, Z-0897-2024, Z-0892-2024, Z-0891-2024, Z-0893-2024, Z-0917-2024, Z-0869-2024, Z-0870-2024 and 46 more

Ongoing Class II Medical devices

Recall started
3 May 2023
Reason
The Exeter V40 Trial Head, catalog number 6364-8-026 (size 26mm, -3) may appear green, when it is meant to be a blue color. Green signifies a size 26mm, +3 Trial Head.
Distribution
International distribution in the country of Ireland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1921-2023

Ongoing Class II Medical devices

Recall started
13 April 2023
Reason
Mislabeled:Labeled Catalog Number: 5531-G-312-E, TRIATHLON X3 TIBIAL BEARING INSERT CS #3 12MM, may contain Catalog Number: 5532-G-413-E, TRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM, and vice versa
Distribution
US Nationwide - Worldwide Distribution: CT, FL, IN, KY, MT, NJ, OR, PA, SC Foreign: China, Japan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1654-2023, Z-1653-2023

Ongoing Class II Medical devices

Recall started
10 March 2023
Reason
Certain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
Distribution
U.S., Netherlands, Japan, China, India, Korea, and Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1386-2023, Z-1385-2023

Ongoing Class III Medical devices

Recall started
7 November 2022
Reason
Instructions for Use (IFU) included in the packaging is missing a translation in the Turkish language. Although the Turkish translation is missing, all content within the IFUs is accurate.
Distribution
International distribution in the country of Turkey.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0736-2023, Z-0737-2023

Ongoing Class II Medical devices

Recall started
21 October 2022
Reason
Simplex HV With Gentamicin CE, 10 PACKS (P/N 6193-1-010) were distributed to customers who had ordered Simplex HV With Gentamicin in US, 10 PACKS (P/N 6195-1-010)
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0240-2023

Ongoing Class II Medical devices

Recall started
24 August 2022
Reason
A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the MK3 tibial bearing. The surgeon needed to complete the surgery by re-assembling and re-inserting the extracted MK3 tibial bearing.
Distribution
Distributed to one customer located in New York.
Initiated
Voluntary: Firm initiated
Recall number
Z-1778-2022

Ongoing Class II Medical devices

Recall started
30 March 2022
Reason
Stryker has discovered a potential product mix where the size of the Triathlon Tritanium Tibial Component inside the package does not match the package labeling.
Distribution
Worldwide distribution - US Nationwide in the states of AL, CA, CO, ID, IN, KS, MI, NV, PA, TN, WA, and WV. The countries of UK and Poland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1237-2022

Ongoing Class II Medical devices

Recall started
14 January 2022
Reason
Potential product mix where the size and/or offset of the Biolox delta Ceramic V40 Femoral Head inside the package does not match the package labeling
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, MD, MN, NC, NJ, NY, OH, OR, PA, TN, TX and the countries of Canada, China, Japan, Spain, UK, Netherlands, Germany & Belgium. ***Updated 3/17/22: U.S., Canada, China, Japan, UK, and Netherlands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0843-2022, Z-0842-2022

Terminated Class II Medical devices

Recall started
30 December 2021
Terminated
14 March 2024
Reason
There is a potential label mix between certain lots of Exeter V40 Cemented Hip (125mm) Stem and Exeter V40 Cemented Hip (150mm) Stem.
Distribution
The product was distributed to the following US states: MI
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0652-2022, Z-0651-2022

Ongoing Class II Medical devices

Recall started
21 December 2021
Reason
There is a potential for the outer Tyvek lid to de-bond from the sealed package.
Distribution
Worldwide distribution - US Puerto Rico and the countries of Canada, and Malaysia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0644-2022, Z-0645-2022

Ongoing Class II Medical devices

Recall started
24 November 2021
Reason
Triathlon Pro Tibial Preparation Tray may have been kitted with the Triathlon Cemented #1-3 Keel Punch (P/N: 6541-2-013) instead of the Triathlon Pro #0-3 Cemented Keel Punch (P/N: 6541-2-013Y) instrument can lead to delay in surgery or if used could lead to unintended bone preparation
Distribution
US Nationwide distribution in the states of GA, MI, NC, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0459-2022

Ongoing Class II Medical devices

Recall started
15 October 2021
Reason
There is a potential for the outer white Tyvek lid to debond from the sealed package.
Distribution
Puerto Rico
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0314-2022, Z-0313-2022

Terminated Class II Medical devices

Recall started
1 July 2021
Terminated
17 May 2023
Reason
The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.
Distribution
Worldwide distribution - US Nationwide distribution in the states AR, CO, MN, MS, NJ, OK, PA, TX and the countries of China and Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-2329-2021

Terminated Class II Medical devices

Recall started
5 March 2021
Terminated
10 May 2024
Reason
The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.
Distribution
Distribution in US to: CA,, GA, NC, NJ, NV, OK, TX, and to Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1379-2021

Terminated Class II Medical devices

Recall started
10 February 2021
Terminated
30 May 2023
Reason
Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.
Distribution
Domestic: AZ, GA, MA, MI, MN, NH, NJ, OR, PA, SC, UT, WA, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1186-2021

Terminated Class II Medical devices

Recall started
16 April 2020
Terminated
15 November 2021
Reason
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Distribution
Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New Hampshire, Illinois, New Jersey, Virginia, Ut
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2085-2020, Z-2094-2020, Z-2088-2020, Z-2084-2020, Z-2091-2020, Z-2093-2020, Z-2090-2020, Z-2086-2020, Z-2092-2020, Z-2096-2020, Z-2087-2020, Z-2089-2020 and 4 more

Terminated Class II Medical devices

Recall started
1 April 2020
Terminated
27 August 2020
Reason
Shipments were transported outside of the required refrigerated temperature conditions. Potential hazards if used include nerve damage resulting in paralysis and increased potential for embolism.
Distribution
US distribution to California Colorado Florida Georgia Illinois Indiana Kansas Michigan Ohio Pennsylvania Texas Utah Virginia Washington Wisconsin
Initiated
Voluntary: Firm initiated
Recall number
Z-1900-2020

Terminated Class II Medical devices

Recall started
7 November 2019
Terminated
31 August 2020
Reason
May have a missing locking wire causing delay in surgery while retrieving a back-up device or implantation of a device that is missing a locking wire
Distribution
CA, FL, GA , IA, MA, MI, NC, OH, WI Foreign: Canada, Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-0811-2020

Terminated Class II Medical devices

Recall started
3 October 2019
Terminated
21 April 2022
Reason
Boxes of the 6-Petal X-Change Wall Mesh Small (Part # 0942-8-025) contain the 4-Petal X-Change Wall Mesh Small (Part # 0942-8-015)
Distribution
Foreign: Australia, France, Finland, Germany, Italy, Japan, New Zealand, Netherlands, Norway, Romania, Spain, Sweden and UK
Initiated
Voluntary: Firm initiated
Recall number
Z-0514-2020

Terminated Class II Medical devices

Recall started
19 August 2019
Terminated
6 May 2020
Reason
Due to device overhang issue that protrudes beyond the medial periphery of the mating femoral component
Distribution
Worldwide distribution - US Nationwide distribution including the states of AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and countries of Argentina,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2760-2019, Z-2759-2019, Z-2757-2019, Z-2758-2019

Terminated Class II Medical devices

Recall started
18 June 2019
Terminated
22 May 2020
Reason
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Distribution
US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-2160-2019

Terminated Class II Medical devices

Recall started
28 November 2018
Terminated
28 May 2020
Reason
The surgical technique is being updated to caution against misuse due to reports of cage fractures occurring intra-operatively and post-operatively.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0605-2019

Terminated Class II Medical devices

Recall started
26 November 2018
Terminated
26 May 2020
Reason
Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.
Distribution
Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-0657-2019

Terminated Class II Medical devices

Recall started
15 November 2018
Terminated
18 June 2020
Reason
Incorrect expiry date of 17-Dec-19 Correct expiry date is 11-Aug-19
Distribution
CO, FL, GA, IA, MA, ME, MI, MN, NC, ND, NH, NY, PA, SC, TN, TX, VA, WI & WV
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0677-2019, Z-0678-2019

Terminated Class II Medical devices

Recall started
12 September 2018
Terminated
16 July 2020
Reason
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
Distribution
US Nationwide Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0207-2019

Terminated Class II Medical devices

Recall started
10 September 2018
Terminated
31 August 2020
Reason
Potential for outer pouch sterile barrier to be compromised
Distribution
US Nationwide in the states of AL, FL, IL, MI, NJ & OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-0373-2019

Terminated Class II Medical devices

Recall started
22 May 2018
Terminated
8 May 2020
Reason
The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.
Distribution
Distributed nationwide. Foreign distribution to Australia, Canada, Chile, China, Brazil, France, Germany, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2293-2018, Z-2292-2018, Z-2294-2018, Z-2298-2018, Z-2297-2018, Z-2296-2018, Z-2299-2018, Z-2295-2018

Terminated Class II Medical devices

Recall started
30 January 2018
Terminated
7 May 2020
Reason
Customers may receive the incorrect size implant from what is labeled on the box.
Distribution
USA (nationwide) Distribution to the states of : AL, AZ, CA, FL, GA, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA
Initiated
Voluntary: Firm initiated
Recall number
Z-1462-2018

Terminated Class II Medical devices

Recall started
12 January 2018
Terminated
7 May 2020
Reason
Firm became aware that the specific product/lot combinations referenced above may contain the incorrect size implant from what is labeled on the box.
Distribution
Affected products were sent to three medical facilities. Affected product will be returned to the manufacturer for disposition.
Initiated
Voluntary: Firm initiated
Recall number
Z-1512-2018

Terminated Class II Medical devices

Recall started
21 December 2017
Terminated
5 October 2018
Reason
From March 2007 to October 2014, product did not contain updated instructions for use (IFU) to clarify the function of locking mechanism and warnings related to the locking screw.
Distribution
NY, FL, KS
Initiated
Voluntary: Firm initiated
Recall number
Z-2567-2018

Terminated Class II Medical devices

Recall started
24 October 2017
Terminated
18 May 2023
Reason
Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Distribution
Distribution US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2729-2018, Z-2694-2018, Z-2676-2018, Z-2715-2018, Z-2713-2018, Z-2668-2018, Z-2674-2018, Z-2688-2018, Z-2725-2018, Z-2675-2018, Z-2672-2018, Z-2716-2018 and 58 more

Terminated Class II Medical devices

Recall started
8 September 2017
Terminated
31 July 2018
Reason
When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall number
Z-0810-2018

Terminated Class II Medical devices

Recall started
21 July 2017
Terminated
12 March 2020
Reason
Inner and outer sterile barriers not fully sealed.
Distribution
Distributed in ten (10) states: AR, DE, FL, IL, IN, MI, NJ, OH, TN, and TX; and Canada, Sweden, United Kingdom, France, Spain, Italy, Japan, Romania, Colombia, Netherlands, and Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0940-2018, Z-0941-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.