Horiba Instruments Incorporated recalls
2 recalls recorded since 2018, most recently on 28 March 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 28 March 2024
- Reason
- HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
- Distribution
- US Nationwide and Worldwide Distribution
- Initiated
- FDA Mandated
- Recall number
- Z-2166-2024
- HORIBA custom configured fluorescence instrument, modular Fluorolog-QM
Terminated Class II Medical devices
- Recall started
- 30 October 2018
- Terminated
- 16 June 2020
- Reason
- Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).
- Distribution
- US: AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0119-2020
- Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.