Horiba Instruments, Inc dba Horiba Medical recalls

2 recalls recorded since 2018, most recently on 26 April 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
26 April 2018
Terminated
26 September 2019
Reason
When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
Distribution
Worldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-2286-2018

Terminated Class II Medical devices

Recall started
21 November 2017
Terminated
4 November 2019
Reason
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0493-2018, Z-0494-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.