Horiba Instruments, Inc dba Horiba Medical recalls
2 recalls recorded since 2018, most recently on 26 April 2018. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 26 April 2018
- Terminated
- 26 September 2019
- Reason
- When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
- Distribution
- Worldwide Distribution: US (nationwide) and to countries of: Bulgaria, Burma, Cameroon, Djibouti, France, Gabon, Germany, Italy, Lebanon, Mali, Mauritania, Poland, Portugal, Romania, Saudi Arabia, South Africa and Thailand.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2286-2018
- ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Terminated Class II Medical devices
- Recall started
- 21 November 2017
- Terminated
- 4 November 2019
- Reason
- The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
- Distribution
- Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0493-2018, Z-0494-2018
- 10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
- 15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400