Hobbs Medical, Inc. recalls

5 recalls recorded since 2018, most recently on 4 October 2024. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)4
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 October 2024
Reason
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Distribution
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0476-2025, Z-0475-2025, Z-0472-2025, Z-0474-2025, Z-0471-2025, Z-0473-2025, Z-0469-2025, Z-0468-2025, Z-0470-2025

Ongoing Class II Medical devices

Recall started
29 July 2022
Reason
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Austria, Belgium, Finland, Norway.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0070-2023, Z-0051-2023, Z-0049-2023, Z-0018-2023, Z-0025-2023, Z-0034-2023, Z-0069-2023, Z-0054-2023, Z-0041-2023, Z-0013-2023, Z-0055-2023, Z-0030-2023 and 49 more

Ongoing Class II Medical devices

Recall started
3 June 2022
Reason
During use, the sheath of the device may accordion at the proximal end where the sheath meets the shrink band. Kinking of the sheath may prevent the snare from fully closing, and injury may result.
Distribution
Domestic distribution to NC, TX, and WA. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1388-2022

Ongoing Class II Medical devices

Recall started
28 December 2021
Reason
The labeling for the Microbiology Brush identifies the device as a Bronchial Microbiology Brush. The device is only cleared for gastrointestinal use and is not cleared for bronchial use.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0711-2022

Terminated Class II Medical devices

Recall started
1 May 2018
Terminated
10 April 2020
Reason
Device was marketed for uses outside the 510(k) clearance.
Distribution
US Nationwide Distribution to states of: AZ, CA, GA, IL, MD, NY, TN, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2052-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.