Hitachi Medical Systems America Inc recalls

6 recalls recorded since 2018, most recently on 1 March 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)0
Terminated or completed6
Class I (FDA)0

Terminated Class II Medical devices

Recall started
1 March 2021
Terminated
4 November 2021
Reason
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Distribution
US Nationwide distribution in the state of TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1282-2021

Terminated Class II Medical devices

Recall started
31 January 2020
Terminated
31 July 2020
Reason
These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
Distribution
Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1341-2020

Terminated Class II Medical devices

Recall started
16 October 2019
Terminated
30 April 2020
Reason
There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
Distribution
The devices were distributed to the following US states: CA, FL, GA, IA, IN, MD, MT, NC, NY, OH, OK, SD, TN, TX, WY, and PR. The products were distributed to the following foreign countries: Brazil and Mexico.
Initiated
Voluntary: Firm initiated
Recall number
Z-0271-2020

Terminated Class II Medical devices

Recall started
29 June 2018
Terminated
27 April 2020
Reason
The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
Distribution
US Distribution to one customer in TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2839-2018

Terminated Class II Medical devices

Recall started
18 June 2018
Terminated
20 April 2020
Reason
The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.
Distribution
Worldwide Distribution - US Nationwide distribution.in the states of Mexico.
Initiated
Voluntary: Firm initiated
Recall number
Z-2479-2018

Terminated Class II Medical devices

Recall started
20 October 2017
Terminated
16 April 2020
Reason
The ultrasound probe may not have adequate protection against electrical shock hazards.
Distribution
Distributed in 21 states: AR, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MO, NC, NY, OK, OR, PA, SC, TX, VA, WA, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-0816-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.