Helena Laboratories, Corp. recalls

14 recalls recorded since 2018, most recently on 2 May 2023. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls14
Ongoing (FDA)6
Terminated or completed8
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
2 May 2023
Reason
Packaging for a serrated blade applicator kit (12 Sample) may contain a serrated blade applicator (18 sample). This results in a characteristic pattern of missed or light samples on the gel, leading to incorrect or no result, or delay in result.
Distribution
Distribution to US states of PA, NE, MI, FL, AR, NY, TX, and CA, and to Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1975-2023

Ongoing Class II Medical devices

Recall started
6 April 2022
Reason
Due to microbial growth causing interference with interpretation.
Distribution
U.S.: FL, MO, OH, and TX O.U.S.: Canada, Uruguay, and Vietnam,
Initiated
Voluntary: Firm initiated
Recall number
Z-1244-2022

Terminated Class II Medical devices

Recall started
5 January 2021
Terminated
19 November 2024
Reason
Intermittent potential for shorter than expected clot times.
Distribution
Worldwide distribution - U.S. Nationwide distribution and the countries of Germany, Italy, Romania, Australia, Chile.
Initiated
Voluntary: Firm initiated
Recall number
Z-1108-2021

Ongoing Class II Medical devices

Recall started
9 November 2020
Reason
Due to complaints received associated with cracked/split test tube caps.
Distribution
Worldwide distribution - U.S. Nationwide including in the states of NJ, NY, PA and TX. The countries of Austria, Canada, England, France, Italy, Portugal, Russia, Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-0635-2021

Ongoing Class II Medical devices

Recall started
8 August 2018
Reason
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
Distribution
Distribution was made to AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WY. There was no foreign/military/government distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1950-2019, Z-1949-2019

Ongoing Class III Medical devices

Recall started
23 May 2018
Reason
Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit may contain particulate matter or microbial growth.
Distribution
US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0536-2020, Z-0537-2020

Ongoing Class III Medical devices

Recall started
27 February 2018
Reason
Lots of gel kits contained a newer revision of the Customer Procedures instructions than intended. The incorrect inserts refer to updated parameters associated with new electrodes -- which have not yet been made available to users.
Distribution
US: OH, NY, MI, IN, IL, MS, TX, CT, NJ, KY, FL, MD, GA, PA, AZ, MO, VA, SC, LA, KS, WA, WI, OUS: Colombia, Brasil, Taiwan
Initiated
Voluntary: Firm initiated
Recall number
Z-0297-2020

Terminated Class II Medical devices

Recall started
7 December 2017
Terminated
9 June 2020
Reason
Some of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the screening assay.
Distribution
Distribution was made to Germany and Australia. There was no U.S. distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2761-2019

Terminated Class II Medical devices

Recall started
21 July 2017
Terminated
19 June 2020
Reason
The plates do not have wells punched for the controls in the patient 1 area.
Distribution
Distribution was made to IN, KY, MA, OH, OR, TN, and TX. There was government distribution. Foreign distribution was made to Chile.
Initiated
Voluntary: Firm initiated
Recall number
Z-0045-2020

Terminated Class II Medical devices

Recall started
4 April 2017
Terminated
14 May 2020
Reason
Internal components of the Protein S ELISA Antigen Kit (Lot: 3-16-5292) were mislabeled. The Protein S Conjugate Solution is labeled as Substrate, and the Substrate is labeled as Protein S conjugate.
Distribution
International distribution to country of: Jordan
Initiated
Voluntary: Firm initiated
Recall number
Z-0106-2019

Terminated Class II Medical devices

Recall started
24 February 2017
Terminated
20 October 2023
Reason
Due to some cards in this lot reported to have experienced variations in results, (verified by wave form), resulting in some cards with shorter or longer times than expected.
Distribution
US: Products under control of Manufacturer in TX OUS: Four International Distributors - Australia, Netherlands, Italy, and England
Initiated
Voluntary: Firm initiated
Recall number
Z-0561-2019

Terminated Class III Medical devices

Recall started
7 November 2016
Terminated
15 January 2021
Reason
Due to a customer experiencing complaints on bone fraction of assay not being in range. The complaint: ~12% lower enzyme activity was noted.
Distribution
Worldwide Distribution: US (nationwide) to states of: GA, IN, and MI, and countries of: Australia, Brazil, Canada, Japan and Hong Kong.
Initiated
Voluntary: Firm initiated
Recall number
Z-0529-2019

Terminated Class II Medical devices

Recall started
14 October 2016
Terminated
12 October 2023
Reason
Due to customer-experienced complaints on narrow assay range compared to prior lots and the user being out of assay range.
Distribution
Domestic: FL, IL, MN, MT, NY, OH, TN, TX, VA, and WI OUS: Australia, Canada, Jordan, and South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-2434-2019

Terminated Class II Medical devices

Recall started
20 April 2016
Terminated
15 February 2024
Reason
Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected
Distribution
US Distribution to state of TX and internationally to: Australia, Netherlands, Italy, and England.
Initiated
Voluntary: Firm initiated
Recall number
Z-0594-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.