Miami Lakes, FL, United States

Heartware, Inc. recalls

19 recalls recorded since 2018, most recently on 31 October 2023. 16 are listed as ongoing. Status and classification are as published by the FDA.

Recalls19
Ongoing (FDA)16
Terminated or completed3
Class I (FDA)8

Ongoing Class II Medical devices

Recall started
31 October 2023
Reason
Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0478-2024, Z-0482-2024, Z-0481-2024, Z-0475-2024, Z-0479-2024, Z-0483-2024, Z-0476-2024, Z-0477-2024, Z-0480-2024

Ongoing Class II Medical devices

Recall started
16 May 2023
Reason
Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal.
Distribution
Worldwide - US Nationwide distribution in the states of TX, MO, VA, and the countries of Bahrain, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-1993-2023

Ongoing Class II Medical devices

Recall started
29 November 2022
Reason
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0742-2023, Z-0747-2023, Z-0743-2023, Z-0745-2023, Z-0746-2023, Z-0750-2023, Z-0744-2023, Z-0749-2023, Z-0748-2023

Ongoing Class I Medical devices

Recall started
28 June 2022
Reason
Battery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery charger, and/or result in the battery to appear to remain charged while discharging.
Distribution
Worldwide distribution. US nationwide, Japan, Indonesia, Brazil, Chile, Belgium, Italy, Germany, Switzerland, United Kingdom, Poland, Hungary, Czech Republic, Mauritius, France, South Africa, Greece, Netherlands, Sweden, Norway, Denmark, Estonia, Lebanon, Croatia, Spain, Austria, United Arab Emirat
Initiated
Voluntary: Firm initiated
Recall number
Z-1512-2022

Ongoing Class II Medical devices

Recall started
30 March 2022
Reason
Finnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Distribution
Finland and Turkey
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1103-2022, Z-1106-2022, Z-1104-2022, Z-1105-2022, Z-1107-2022, Z-1102-2022

Ongoing Class II Medical devices

Recall started
27 January 2022
Reason
Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.
Distribution
Distribution throughout the United States. OUS distribution in Argentina, Armenia, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea (So
Initiated
Voluntary: Firm initiated
Recall number
Z-0726-2022

Ongoing Class II Medical devices

Recall started
6 August 2021
Reason
There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.
Distribution
Distribution throughout the US. Also distribution in ARGENTINA Armenia, Australia, Austria, BELGIUM, BRAZIL, CANADA, CHILE, Colombia, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, Germany, GREECE HONG KONG, HUNGARY, INDIA, ISRAEL, Italy, JAPAN, Kazakhstan, Korea, KUWAIT, La
Initiated
Voluntary: Firm initiated
Recall number
Z-2478-2021

Ongoing Class I Medical devices

Recall started
3 June 2021
Reason
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.
Distribution
Worldwide distribution. US nationwide and Armenia, AUSTRIA, Bahrain, Belgium, CROATIA, CZECH REPUBLIC, Denmark, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, Hungary, Israel, ITALY, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, LITHUANIA, LUXEMBOURG, Macedonia, NETHERLANDS, Norway, POLAND, Qatar
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2190-2021, Z-2191-2021, Z-2196-2021, Z-2197-2021, Z-2192-2021, Z-2188-2021, Z-2189-2021, Z-2194-2021, Z-2193-2021, Z-2195-2021

Terminated Class II Medical devices

Recall started
28 May 2021
Terminated
22 September 2026
Reason
The pump has an impeller with a shroud height that did not meet the lower control limit.
Distribution
International distribution to the country of Germany.
Initiated
Voluntary: Firm initiated
Recall number
Z-2072-2021

Ongoing Class II Medical devices

Recall started
26 February 2021
Reason
IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Italy, Great Britain, Germany, Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1438-2021

Ongoing Class II Medical devices

Recall started
26 February 2021
Reason
IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604
Distribution
Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-1469-2021

Ongoing Class I Medical devices

Recall started
26 February 2021
Reason
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1336-2021, Z-1340-2021, Z-1339-2021, Z-1337-2021, Z-1338-2021, Z-1341-2021

Ongoing Class I Medical devices

Recall started
26 February 2021
Reason
Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the support strap and wear instructions in addition to cleaning instructions of the carrying cases. The IFU and PM are being updated to add a useful life for the carrying cases. The PM is being updated to inform users to keep the driveline cover on when disconnecting and reconnecting the driveline. The IFU and PM are being updated to clarify the power-up sequence that causes the alarm indicator LEDs and both sets of battery LEDs to turn red for 2.5 seconds while the LCD displays the power-on message. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87603 and 87604
Distribution
Worldwide distribution: US (nationwide) and countries of: ARGENTINA, Armenia, Australia, AUSTRIA, Bahrain, BELGIUM, Brazil, CANADA, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, FINLAND, France, GERMANY, Greece, Hong Kong, Hungary, INDIA, Israel, Italy, Japan, Kazakhstan, Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-1426-2021

Ongoing Class I Medical devices

Recall started
19 November 2020
Reason
Pumps manufactured with impellers from a subset of lots from a single supplier are failing to initially start, restart, or experiencing a delay in restart within the maximum number of restart attempts at a rate substantially higher than pump in the overall population.
Distribution
Worldwide - US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0946-2021

Ongoing Class I Medical devices

Recall started
3 April 2020
Reason
The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2055-2020, Z-2058-2020, Z-2056-2020, Z-2057-2020, Z-2059-2020

Ongoing Class I Medical devices

Recall started
23 January 2020
Reason
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
Distribution
No affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic o
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1427-2020, Z-1428-2020

Terminated Class II Medical devices

Recall started
19 November 2018
Terminated
21 July 2020
Reason
HVAD Battery Charger units manufactured with wrong inductors.
Distribution
Worldwide distribution - US nationwide distribution in the state of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI. Countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Estonia, Fi
Initiated
Voluntary: Firm initiated
Recall number
Z-0582-2019

Terminated Class II Medical devices

Recall started
28 August 2018
Terminated
5 April 2019
Reason
Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0067-2019

Ongoing Class I Medical devices

Recall started
2 May 2018
Reason
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.
Distribution
Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Le
Initiated
Voluntary: Firm initiated
Recall number
Z-1903-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.