HeartSine Technologies Ltd recalls

8 recalls recorded since 2018, most recently on 27 July 2026. 8 are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)8
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
27 July 2026
Reason
Due to a damaged (over stretched) membrane tail that causes an intermittent fault.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME MI, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SD,TN, TX, UT, VA, WA, WI, WV and WY. The countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Hong Ko
Initiated
Voluntary: Firm initiated
Recall number
Z-3120-2026

Ongoing Class III Medical devices

Recall started
27 July 2026
Reason
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
Distribution
U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-3270-2026

Ongoing Class II Medical devices

Recall started
18 September 2025
Reason
Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
Distribution
US Nationwide distribution. Worldwide distribution to Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Ita
Initiated
Voluntary: Firm initiated
Recall number
Z-0288-2026

Ongoing Class II Medical devices

Recall started
30 June 2025
Reason
Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina,
Initiated
Voluntary: Firm initiated
Recall number
Z-2174-2025

Ongoing Class II Medical devices

Recall started
8 April 2024
Reason
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GC, HI, IA, ID, IL, IN, KS, KY, LA, MA, MB, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-1820-2024

Ongoing Class II Medical devices

Recall started
14 February 2024
Reason
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
Distribution
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1141-2024

Ongoing Class II Medical devices

Recall started
26 April 2023
Reason
Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or delivering therapy correctly.
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA & WI and OUS (countries of): Australia, Canada, Denmark, France, Iceland, Ireland, Israel, Italy, Japan, Korea,
Initiated
Voluntary: Firm initiated
Recall number
Z-2143-2023

Ongoing Class II Medical devices

Recall started
1 February 2023
Reason
Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in therapy or no therapy.
Distribution
US: NJ, PA, WI, NE, TX, PA, CA, AZ, AR OUS: Australia, Czechia, Canada, New Zealand
Initiated
Voluntary: Firm initiated
Recall number
Z-1217-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.