Recall Record

Reno, NV, United States

Hamilton Medical, Inc. recalls

3 recalls recorded since 2018, most recently on 26 July 2023. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)2

Ongoing Class I Medical devices

Recall started
26 July 2023
Reason
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2529-2023, Z-2526-2023, Z-2528-2023, Z-2525-2023, Z-2530-2023, Z-2527-2023

Ongoing Class I Medical devices

Recall started
15 June 2023
Reason
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Distribution
U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New York North Carolina No
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2142-2023, Z-2140-2023, Z-2141-2023, Z-2139-2023

Ongoing Class II Medical devices

Recall started
20 June 2017
Reason
Issue related to the performance of the motor in recalled product. During use, motor may cease to function.
Distribution
US Distribution to IL and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0335-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.