12 recalls recorded since 2018, most recently on 29 May 2026. 9 are listed as ongoing. Status and classification are as published by the FDA.

Recalls12
Ongoing (FDA)9
Terminated or completed3
Class I (FDA)6

Ongoing Class I Medical devices

Recall started
29 May 2026
Reason
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2463-2026

Ongoing Class II Medical devices

Recall started
22 May 2026
Reason
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2625-2026

Ongoing Class I Medical devices

Recall started
26 June 2025
Reason
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
Distribution
US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.
Initiated
Voluntary: Firm initiated
Recall number
Z-2167-2025

Ongoing Class I Medical devices

Recall started
3 June 2025
Reason
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2328-2025

Ongoing Class I Medical devices

Recall started
15 May 2024
Reason
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Distribution
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2020-2024

Ongoing Class II Medical devices

Recall started
14 March 2023
Reason
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.
Distribution
US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1429-2023

Ongoing Class II Medical devices

Recall started
22 September 2022
Reason
Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.
Distribution
U.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN, KS, KY, MA, MI, MO, NV, NM, NY, NC, OH, PA, Puerto Rico, SC, TX, VA, WA, WV, WI, and WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-0267-2023

Ongoing Class I Medical devices

Recall started
27 June 2022
Reason
Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed
Distribution
US nationwide distribution including Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-1525-2022

Completed Class III Medical devices

Recall started
5 January 2021
Reason
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.
Distribution
US: WI, FL, CA, NV, PA, TX, PA
Initiated
Voluntary: Firm initiated
Recall number
Z-1385-2022

Terminated Class I Medical devices

Recall started
25 March 2019
Terminated
3 June 2024
Reason
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.
Distribution
Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT, and RI. And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE,
Initiated
Voluntary: Firm initiated
Recall number
Z-1874-2019

Ongoing Class II Medical devices

Recall started
12 March 2018
Reason
After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
Distribution
Worldwide Distribution -- USA, including the states of AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MI, MN, MO, MT, NC, ND, NE, NV, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Algeria, Angola, Argentina, Armenia, Australia, Austria, Austria, Azerbaijan, Bangladesh, Belarus, Belg
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1371-2018, Z-1372-2018

Terminated Class II Medical devices

Recall started
9 March 2018
Terminated
18 March 2019
Reason
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Distribution
US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2796-2018, Z-2795-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.