Haemonetics Corporation recalls

6 recalls recorded since 2018, most recently on 26 April 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)0
Terminated or completed6
Class I (FDA)0

Terminated Class II Medical devices

Recall started
26 April 2022
Terminated
15 March 2024
Reason
Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference range, it could lead to misdiagnosis and incorrect treatment or failure to treat.
Distribution
Domestic distribution US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1310-2022

Terminated Class II Medical devices

Recall started
19 April 2022
Terminated
27 March 2024
Reason
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1298-2022

Terminated Class II Medical devices

Recall started
19 February 2020
Terminated
21 June 2021
Reason
Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1617-2020

Terminated Class II Medical devices

Recall started
21 October 2019
Terminated
6 November 2020
Reason
Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0904-2020, Z-0890-2020, Z-0900-2020, Z-0895-2020, Z-0906-2020, Z-0889-2020, Z-0896-2020, Z-0891-2020, Z-0899-2020, Z-0901-2020, Z-0892-2020, Z-0894-2020 and 10 more

Terminated Class II Medical devices

Recall started
13 July 2018
Terminated
16 June 2020
Reason
OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2950-2018, Z-2951-2018

Terminated Class II Medical devices

Recall started
2 August 2017
Terminated
6 August 2020
Reason
Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
Distribution
US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0779-2019, Z-0778-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.