Graphic Controls Acquisition Corporation recalls
1 recall recorded since 2018, most recently on 15 March 2022. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 15 March 2022
- Terminated
- 12 July 2024
- Reason
- Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
- Distribution
- US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0936-2022, Z-0938-2022, Z-0940-2022, Z-0937-2022, Z-0939-2022, Z-0935-2022
- CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US
- CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US
- CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model: 32028781 50604-US
- CLARAVUE Pre-wired Electrode - IP SET M5 RT ADULT-US Model: 32028778 50601-US
- CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US
- CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.