Recall Record

Ballerup, Denmark

GN Hearing A/S recalls

1 recall recorded since 2018, most recently on 18 September 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
18 September 2019
Terminated
31 May 2023
Reason
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0584-2020, Z-0583-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.