Globus Medical, Inc. recalls

5 recalls recorded since 2018, most recently on 3 April 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)4
Terminated or completed1
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
3 April 2025
Reason
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
Distribution
US Nationwide distribution in the state of AR.
Initiated
Voluntary: Firm initiated
Recall number
Z-1657-2025

Ongoing Class II Medical devices

Recall started
15 December 2023
Reason
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Distribution
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1007-2024

Ongoing Class II Medical devices

Recall started
11 December 2023
Reason
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
Distribution
Worldwide - US Nationwide and the countries of Czech Republic, Poland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0997-2024, Z-0995-2024, Z-0996-2024

Ongoing Class I Medical devices

Recall started
7 December 2023
Reason
A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
Distribution
Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-0642-2024

Terminated Class II Medical devices

Recall started
22 July 2020
Terminated
19 August 2021
Reason
Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2856-2020, Z-2861-2020, Z-2871-2020, Z-2865-2020, Z-2868-2020, Z-2869-2020, Z-2870-2020, Z-2862-2020, Z-2854-2020, Z-2858-2020, Z-2855-2020, Z-2859-2020 and 10 more

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.